[Contact surfaces, compressive forces and pressure distribution in osteosynthesis. Model studies using pressure-sensitive foil].

Abstract

Previous pressure gauges have been unable to measure exactly the stress over the cross-section of fractures or osteotomies. This first became possible with the pressure sensitive foil. Known operative procedures were examined with this foil using a self-compressing plate and a synthetic tube model. Contact area, compressive force, pressure and pressure distribution were measured simultaneously and analogously. A straight plate produces extremely little force and pressure on the gap. Both are limited to about a quarter of the cross-section just below the plate. The consequence is no sufficient stability. Also, using a prebent plate results in inhomogeneous pressure distribution with high values, but a distinct decrease of the stress on the opposite side of the plate, especially by compression applied at the end of the plate. But compression from near the fracture side gives the highest values and a most homogeneous force distribution. In this case the force recorded 1700 N, the mean pressure 5,2 MPa and the contact 90% of the cross-section. The length thrust of the plate ends approximately at a torque of screw tightening at 1,0 Nm. Using torques up to 2,5 Nm, force and pressure increase isometrically. Higher torques have no influence. Converging screws result in smaller contact, force and pressure, because they produce a vector force away from the gap and a smaller length thrust.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

H J Hehne, U Soltész, R Desiderato. [Contact surfaces, compressive forces and pressure distribution in osteosynthesis. Model studies using pressure-sensitive foil].. https://doi.org/10.1055/s-2008-1039716

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

The effect of excessive tibial torsion on the capacity of muscles to extend the hip and knee during single-limb stance.

Excessive tibial torsion, a rotational deformity about the long axis of the tibia, is common in patients with cerebral palsy who walk with a crouch gait. Previous research suggests that this deformity may contribute to crouch gait by reducing the capacity of soleus to extend the knee; however, the effects of excess external torsion on the capacity of other muscles to extend the stance limb during walking are unknown. A computer model of the musculoskeletal system was developed to simulate a range of tibial torsion deformities. A dynamic analysis was then performed to determine the effect of these deformities on the capacity of lower limb muscles to extend the hip and knee at body positions corresponding to the single-limb stance phase of a normal gait cycle. Analysis of the model confirmed that excessive external torsion reduces the extension capacity of soleus. In addition, our analysis revealed that several important muscles crossing the hip and knee are also adversely affected by excessive tibial torsion. With a tibial torsion deformity of 30 degrees , the capacities of soleus, posterior gluteus medius, and gluteus maximus to extend both the hip and knee were all reduced by over 10%. Since a tibial torsion deformity reduces the capacity of muscles to extend the hip and knee, it may be a significant contributor to crouch gait, especially when greater than 30 degrees from normal, and thus should be considered by clinicians when making treatment decisions.

Biomechanical Phenomena↗

Predicting the subject-specific primary stability of cementless implants during pre-operative planning: preliminary validation of subject-specific finite-element models.

Pre-operative planning help the surgeon in taking the proper clinical decision. The ultimate goal of this work is to develop numerical models that allow the surgeon to estimate the primary stability during the pre-operative planning session. The present study was aimed to validate finite-element (FE) models accounting for patient and prosthetic size and position as planned by the surgeon. For this purpose, the FE model of a cadaveric femur was generated starting from the CT scan and the anatomical position of a cementless stem derived by a skilled surgeon using a pre-operative CT-based planning simulation software. In-vitro experimental measurements were used as benchmark problem to validate the bone-implant relative micromotions predicted by the patient-specific FE model. A maximum torque in internal rotation of 11.4 Nm was applied to the proximal part of the hip stem. The error on the maximum predicted micromotion was 12% of the peak micromotion measured experimentally. The average error over the entire range of applied torques was only 7% of peak measurement. Hence, the present study confirms that it is possible to accurately predict the level of primary stability achieved for cementless stems using numerical models that account for patient specificity and surgical variability.

Biomechanical Phenomena↗

Experimental validation of intact and implanted distal femur finite element models.

Four finite element (FE) models of intact and distal femur of knee replacements were validated relative to measured bone strains. FE models of linear tetrahedrons were used. Femoral replacements with cemented stemless, cemented and noncemented femoral stems of the PFC Sigma Modular Knee System were analyzed. Bone strains were recorded at ten locations on the cortex. The magnitude of the FE bone strains corresponded to the mean measured strains, with an overall agreement of 10%. Linear regression between the FE and mean experimental strains produced slopes between 0.94 and 1.06 and R(2) values between 0.92 and 0.99. RSME values were less than 12%. The FE models were able to adequately replicate the mechanical behavior of distal femur reconstructions.

Biomechanical Phenomena↗