[Comparative echocardiographic volume determinations using a dynamic heart model].

Abstract

Using a dynamic and symmetrical cardiac phantom different echocardiographic mathematical models (Simpson 7 slices, area-length method, Simpson 2 slices and method according to Teichholz) were compared. 9 different end-diastolic (EDV) and end-systolic (ESV) volumes, 9 different stroke volumes (SV) and ejection fractions (EF) were used. EDV and ESV varied between 39-298 ml; SV between 29-100 ml and EF between 14-46%. In addition 10 fixed volumes of the same shape were evaluated using the same echocardiographic mathematical models. While symmetrical fixed volumes can be assessed correctly (r = 0.97-0.98), apart from the formula according to Teichholz (r = 0.89, significant underestimation of volumes), the correlation coefficients decrease using a dynamic cardiac phantom. In the modification of Simpson with 7 slices the best correlation was found for all parameters (EDV: r = 0.93; ESV: r = 0.94; EF: r = 0.87; SV: r = 0.81). The biplane area-length method has no advantages over Simpson's rule with 2 slices in the short axis; for symmetrical models both methods are comparable, both having high correlation coefficients (for volumes r = 0.85 and r = 0.88; for EF 0.78 and 0.84). Using the method according to Teichholz symmetrical volumes can be well assessed (r = 0.90), for the determination of EF the correlation coefficient decreases to r = 0.65 and for stroke volume to 0.33, reflecting no significant correlation to the actual SV. Possible causes for a poorer correlation are discussed for moving objects as opposed to the fixed volumes.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

K Berwing, M Kindler, M Wöllenstein, H Bahawar, G Schumacher, M Schlepper. 1985. [Comparative echocardiographic volume determinations using a dynamic heart model].. https://pubmed.ncbi.nlm.nih.gov/4013462/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

[Impedance cardiography--non-invasive measurement of central hemodynamic data].

The theory, method and apparatus behind a new computer technology for transthoracic electrical bioimpedance are described. Measurements and calculations of changes in the electrical conductivity of the thoracic segment during systolic upstroke provides the basis for non-invasive determination of stroke volume and cardiac output. Bioimpedance further offers a sensitive indicator of the content of thoracic fluid. Reports comparing measurements of cardiac output in various clinical conditions by bioimpedance versus invasive methods in general show good correlation. The field of applications and limitations of bioimpedance are described. Measurements are reproducible and the method can also be used in evaluating the cardiodynamic exercise response.

Cardiac Output