The hearing aid feedback path: mathematical simulations and experimental verification.

Abstract

Acoustic feedback in hearing aids has received little attention in the literature. Feedback occurs when stability conditions of the open-loop transfer function of an in situ hearing aid are violated. Solving the feedback problem will first require knowledge of the open-loop transfer function. Included in the open-loop transfer function is the acoustical path by which sound emanating from the earmold vent returns to the microphone (i.e., the feedback path). Reported herein are two different mathematical procedures for simulating transfer functions of the feedback path of an eyeglass-type hearing aid. In one procedure the vent exit was modeled as a point source of sound located on a flat plane, while it was treated as a point source on a sphere in the other. Results of laboratory experiments indicate that the mathematical models accurately predict those acoustic phenomena for which they were intended: point sources on plane and spherical baffles. Results of manikin experiments showed both models to be less accurate for simulating the feedback path around the human head. The maximum difference between experiment and theory was 6 dB at one frequency. Surprisingly, the flat-baffle model produced better agreement with experimental results than did the sphere model.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

D P Egolf, H C Howell, K A Weaver, D S Barker. 1985. The hearing aid feedback path: mathematical simulations and experimental verification.. https://doi.org/10.1121/1.392795

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Effects of gamma irradiation on mechanical properties of defatted trabecular bone allografts assessed by speed-of-sound measurement.

New sterilization methods for human bone allografts may lead to alterations in bone mechanical properties, which strongly influence short- and medium-term outcomes. In many sterilization procedures, bone allografts are subjected to gamma irradiation, usually with 25 KGy, after treatment and packaging. We used speed-of-sound (SOS) measurements to evaluate the effects of gamma irradiation on bone. All bone specimens were subjected to the same microbial inactivation procedure. They were then separated into three groups, of which one was treated and not irradiated and two were exposed to 10 and 25 KGy of gamma radiation, respectively. SOS was measured using high- and low-frequency ultrasound beams in each orthogonal direction. SOS and Young modulus were altered significantly in the three groups, compared to native untreated bone. Exposure to 10 or 25 KGy had no noticeable effect on the study variables. The impact of irradiation was small compared to the effects of physical or chemical defatting. Reducing the radiation dose used in everyday practice failed to improve graft mechanical properties in this study.

Acoustics↗

Objective and subjective evaluation of the acoustic comfort in classrooms.

The acoustic comfort of classrooms in a Brazilian public school has been evaluated through interviews with 62 teachers and 464 pupils, measurements of background noise, reverberation time, and sound insulation. Acoustic measurements have revealed the poor acoustic quality of the classrooms. Results have shown that teachers and pupils consider the noise generated and the voice of the teacher in neighboring classrooms as the main sources of annoyance inside the classroom. Acoustic simulations resulted in the suggestion of placement of perforated plywood on the ceiling, for reduction in reverberation time and increase in the acoustic comfort of the classrooms.

Acoustics↗

Analysis of effective radiating area, power, intensity, and field characteristics of ultrasound transducers.

OBJECTIVE: To characterize the ultrasound fields produced by a cohort of transducers from a single manufacturer via hydrophone and Schlieren technology. DESIGN: Descriptive study. SETTING: Measurement laboratory. PARTICIPANTS: Seven same-model ultrasound transducers from a single manufacturer. INTERVENTIONS: Not applicable. MAIN OUTCOME MEASURES: Effective radiating area (ERA), total power, spatial average intensity (SAI), beam nonuniformity ratio (BNR), and Schlieren beam widths at 1.0 and 3.3 MHz. RESULTS: Values for ERA (1.0 MHz range, 3.62-4.38 cm(2); 3.3 MHz range, 3.74-4.76 cm(2)), total power (1.0 MHz range, 5.0-5.6 W; 3.3 MHz range, 4.7-5.7 W), SAI (1.0 MHz range, 1.2-1.4 W/cm(2); 3.3 MHz range, 1.0-1.5 W/cm(2)), and BNR (1.0 MHz range, 2.79-5.85; 3.3 MHz range, 2.51-4.56) fell within manufacturer's specifications and U.S. Food and Drug Administration (FDA) regulations. Schlieren analysis showed significantly larger beam widths at 3.3 MHz compared with 1.0 MHz and a large degree of variability in the ultrasound fields generated by the different transducers. There were no significant correlations between beam widths and ERA values. CONCLUSIONS: ERA and total power values in a test cohort exist within a range that met FDA regulations. Individual variability in ERA and total power resulted in 50% variability in SAI. This variability may help explain previous reports of heating differences between transducers.

Acoustics↗