PubMed · 6197284
Penicillin activity in brain tissue: a method for continuous measurement.
Abstract
A double-barrelled microelectrode is described which permits the continuous and simultaneous measurement of penicillin concentration and of the local bioelectric activity in nervous tissue. The electrode, which is based on a chloride ligand (Corning no. 477315), has a short response time and enables reproducible measurements of penicillin concentration even below 500 IU/ml.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
E J Speckmann, C E Elger, A Lehmenkuehler. 1983. Penicillin activity in brain tissue: a method for continuous measurement.. https://doi.org/10.1016/0013-4694(83)90033-0