Pacemaker-mediated tachycardias: an unresolved problem.
Abstract
The advent of dual chamber pacing systems that restore atrioventricular synchrony and allow rate variability in response to physiologic stress has provided a major advance in pacing therapy. Undesirable pacemaker-mediated tachycardias may arise when the atrial sensing amplifier of a VDD or DDD device detects a retrogradely conducted P-wave, ectopic atrial tachycardia, myopectoral contractions, or environmental electromagnetic interference. The rate of the resultant tachycardia is determined by the programmed upper tracking rate of the pacemaker. Although seldom life-threatening, troublesome myocardial ischemia may ensue in patients with coronary artery disease. In most instances, the incidence and rate of pacemaker-mediated tachycardias may be minimized by judicious programming.