PubMed · 6421986
Blood ammonia measurement using a simple reflectometer.
Abstract
We have assessed a compact Blood Ammonia Checker System (Ammonia Checker) consisting of a reflectometer which measures the intensity of colour formed by blood ammonia on a bromocresol green indicator reagent plate. The results show good correlation (r = 0.95) with our routine chemical method and a regression line y = 0.835X - 1.468. The Ammonia Checker has good precision (CV less than 10%) at the critical blood ammonia concentration and accurately measures predetermined ammonia concentrations. Blood sampling is improved with a precision pipette. The Ammonia Checker is simple, convenient and reliable and is ideally suited for a laboratory required to perform an urgent blood ammonia measurement.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
R N Ratnaike, J E Buttery, S Hoffmann. 1984. Blood ammonia measurement using a simple reflectometer.. https://doi.org/10.1515/cclm.1984.22.1.105