Artefactual decrease in total protein concentration in patients with monoclonal gammopathies: a method-dependent error.

Abstract

We found total protein estimates in patients with monoclonal gammopathies to be erroneously low when using the Cobas-Bio centrifugal analyzer. This problem occurred only when a serum-water blank was used. This probably results from the precipitation of these proteins under conditions of low ionic strength resulting in high blank readings. The problem can be avoided if a serum-saline blank is used.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

K J Whitlow, A A Nanji, M R Pudek. 1984. Artefactual decrease in total protein concentration in patients with monoclonal gammopathies: a method-dependent error.. https://doi.org/10.1016/s0009-9120(84)90123-1

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

[Duplication in a medical department of tests previously performed in a primary sector].

The aim of this study was to evaluate whether an insufficient flow of information from the primary to the secondary sector was the cause of unnecessary duplications of investigations. From practitioners of 202 patients admitted to a medical department information about paraclinical examinations during the previous year was obtained. In all 111 general practitioners participated. An average of 3.0 blood analyses and 0.9 other investigations were carried out on each patient. 44% of the patients were not investigated at all. An average of 1.9 blood analyses and 0.2 other investigations on each patient were repeated. 22% of the repetitions in both cases were due to insufficient communication between the practitioner and the hospital. These figures seem low and insufficient communication between the primary and the secondary sector appears to be a minor problem.

Blood Chemical Analysis