PubMed Health⌕ Search

PubMed · 6499413

Finn chamber patch test technique.

Abstract

Proper technique aids reliable patch test results. The Finn chamber volume was determined to be 25 microliters. These studies suggest the optimal volume semi-solid test substance to be 12-15 microliters. More than 18-21 microliters caused significant extrusion of test substance and improper adhesive adherence. Finn chamber paper discs used with solutions absorb a maximum of 17.5 microliters test solutions and keep 15 microliters within the chamber. The evaporation rate of aqueous test solutions from Finn chambers was 1 mg/3 min; thus, the chambers should be applied quickly to avoid drying.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

T Fischer, H Maibach. 1984. Finn chamber patch test technique.. https://doi.org/10.1111/j.1600-0536.1984.tb00957.x

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations