A scanning system for chest radiography with regional exposure control: practical implementation.

Abstract

An experimental scanning apparatus for chest radiography is described which offers good scatter rejection and regional manipulation of film exposure. The apparatus is capable of studying two scanning geometries that uses either a fan of radiation swept laterally over the film in 5s or a spot of radiation scanned over two dimensions in a raster pattern in 8.8s. The manipulation of tube output during the scan is achieved by pulse width modulation with film exposure monitored by a fluorescence detector placed behind the film cassette. Measurement of the scatter rejection properties of the system show that it is superior to that of a 10:1 grid when used with 120-kVp radiation. Phantom images with both geometries demonstrate the capability of the system to improve uniformity of film exposure and soft tissue contrast throughout the mediastinal, diaphragmatic, and retrocardiac areas. The relative characteristics of the one- versus two-dimension scan geometries are given. Clinical raster images formed with the 8.8-s scan time show signs of motion unsharpness which becomes insignificant when the scan time is reduced to 4.5 s.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

D B Plewes, E Vogelstein. A scanning system for chest radiography with regional exposure control: practical implementation.. https://doi.org/10.1118/1.595333

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Scatter fractions in AMBER imaging.

Images of two phantoms were obtained with use of an advanced multiple-beam equalization radiography system, and scatter fractions were estimated with use of a photostimulable phosphor imaging system. Scatter fractions in the equalized images were lower in the mediastinum-equivalent areas and higher in the lung-equivalent areas, relative to images that were conventionally acquired with use of an antiscatter grid. The differences are attributed to a reduction in incident exposure in the lungs and the presence of cross-scatter between lung and mediastinal regions.

Radiography, Thoracic↗