A variable angle slant-hole collimator.

Abstract

A variable-angle slant-hole (VASH) collimator was constructed to show the feasibility of using multiple sliding plates to achieve a range of collimator channel inclinations. One hundred and sixty tungsten plates, 0.125 mm thick and 14 cm square, were photoetched to produce 3025 1.5-mm2 holes in each plate, separated by 0.8-mm septa. Along with the collimator holes, registration holes and positioning grooves were also etched. The plates were placed in a holder and stacked to form a collimator 2.0 cm high. The holder permitted the plates to be "sheared" to achieve viewing angles from 0 to 40 degrees from the vertical. Resolution and sensitivity were determined both across and along the shear directions. Resolution of a thin Tc-99m source, 1.24 mm diam and 7 cm long, located 5 cm from the collimator face in air, was 1.1 cm FWHM at 0 degree shear and remained unchanged with increasing slant. The resolution was similar both across and along the shear plane. Sensitivity was determined with a point source placed 7 cm from the collimator face. At 0 degree slant the sensitivity was 169 cps/MBq (6.24 csp/mu Ci). A general all purpose (GAP) collimator had a FWHM of 1 cm for the line source in air at 5 cm, and a sensitivity of 205 cps/MBq (7.58 cps/mu Ci) for the point source at 7 cm. The data suggest that a variable-angle slant-hole collimator can be constructed of laminated plates.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

R H Moore, N M Alpert, H W Strauss. 1983. A variable angle slant-hole collimator.. https://pubmed.ncbi.nlm.nih.gov/6848704/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Advanced interactive planning techniques for conformal therapy: high level beam descriptions and volumetric mapping techniques.

PURPOSE: To aid in design of conformal radiation therapy treatment plans involving many conformally shaped fields, this work investigates the use of two methodologies to enhance the ease of interactive treatment planning: high-level beam constructs and beam's-eye view volumetric mapping. METHODS AND MATERIALS: High-performance computer graphics running on various workstations using a graphical visualization system (AVS) have been used in this work. Software specific to this application has been written in standard FORTRAN and C languages. A new methodology is introduced by defining radiation therapy "fields" to be composed of multiple beam "segments." Fields can then be defined as higher-level entities such as arcs, cones, and other shapes. A "segmental cone" field, for example, is defined by a symmetry axis and a cone angle, and can be used to rapidly place a series of beam segments that converge at the target volume, while reducing the degree of overlap elsewhere. A new beam's-eye view (BEV) volumetric mapping technique is presented to aid in selecting the placement of conformal radiation fields. With this technique, the relative average dose within an organ of interest is calculated for a sampling of isocentric, conformally shaped beams and displayed either as a "globe," which can be combined with the display of anatomical surfaces, or as a two-dimensionally mapped projection. The dose maps from multiple organs can be generated, stacked, or composited with relative weightings to aid in the placement of fields that minimize overlap with critical structures. RESULTS: The use of these new methodologies is demonstrated for prostate and lung treatment sites and compared to conventional planning techniques. DISCUSSION: The use of many beams for conformal treatment delivery is difficult with current interactive planning. The use of high-level beam constructs provides a means to quickly specify, place, and configure multiple beam arrangements. The BEV volumetrics aids in the placing of fields, which minimize involvement with critical normal tissues. CONCLUSIONS: Early experience with the new methodologies suggest that the new methods help to enhance (or at least speed up) the ability of a treatment planner to create optimal radiation treatment field arrangements.

Evaluation Studies as Topic