PubMed · 6919413
Methicillin-resistant Staphylococcus aureus susceptibility testing with Abbott MS-2 system.
Abstract
The antimicrobial susceptibilities of 100 methicillin=resistant Staphylococcus aureus strains were concurrently determined by the Abbott MS-2 System and by the standard disk diffusion method. Agreement between the two methods was 94% or greater for all of the antibiotics tested except for methicillin and gentamicin. This study indicates that the Abbott MS-2 cannot be relied upon for detection of methicillin resistance in clinical S. aureus isolates.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
J R Carlson, F E Conley, D L Cahall. 1982. Methicillin-resistant Staphylococcus aureus susceptibility testing with Abbott MS-2 system.. https://doi.org/10.1128/aac.21.4.676
Cite the original work for its findings. Save a collection to share your selection of sources.