The oscillotonometer in theory and practice.

Abstract

The accuracy of the Scala Alternans Altera oscillotonometer was investigated by comparison with direct intra-arterial pressures. It was found that the traditional criteria for detecting systolic and diastolic pressures were in error. A slow deflation rate was essential for the accurate measurement of systolic pressure, the maximum amplitude of needle oscillation correlated closely with mean pressure, and the determination of diastolic pressure was so inaccurate as to be clinically useless. The response of the oscillotonometer was also found to be dependent upon the individual combination of oscillotonometer and patient.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

P Hutton, C Prys-Roberts. 1982. The oscillotonometer in theory and practice.. https://doi.org/10.1093/bja%2F54.6.581

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Central venous pulse pressure analysis using an R-synchronized pressure measurement system.

OBJECTIVE: The information derived from central venous catheters is underused. We developed an EKG-R synchronization and averaging system to obtained distinct CVP waveforms and analyzed components of these. METHODS: Twenty-five paralyzed surgical patients undergoing CVP monitoring under mechanical ventilation were studied. CVP and EKG signals were analyzed employing our system, the mean CVP and CVP at end-diastole during expiration were compared, and CVP waveform components were measured using this system. RESULTS: CVP waveforms were clearly visualized in all patients. They showed the a peak to be 1.8+/- 0.7 mmHg, which was the highest of three peaks, and the x trough to be lower than the y trough (-1.6+/- 0.7 mmHg and -0.9+/- 0.5 mmHg, respectively), with a mean pulse pressure of 3.4 mmHg. The difference between the mean CVP and CVP at end-diastole during expiration was 0.58+/- 0.81 mmHg. CONCLUSIONS: The mean CVP can be used as an index of right ventricular preload in patients under mechanical ventilation with regular sinus rhythm. Our newly developed system is useful for clinical monitoring and for education in circulatory physiology.

Blood Pressure Determination↗

Beyond inpatient and outpatient care: alternative model for hypertension management.

Hypertension is a major contributor to worldwide cardiovascular mortality, however, only one-third of patients with hypertension have their blood pressure treated to guideline recommended levels. To improve hypertension control, there may need to be a fundamental shift in care delivery, one that is population-based and simultaneously addresses patient, provider and system barriers. One potential approach is home-based disease management, based on the triad of home monitoring, team care, and patient self-care. Although there may be challenges to achieving the vision of home-based disease management, there are tremendous potential benefits of such an approach for reducing the global burden of cardiovascular disease.

Blood Pressure Determination↗