The continuous thermodilution method for measuring high blood flows.
Abstract
The continuous thermodilution method for the measurement of blood flow from 300 to 1500 ml/min was evaluated in vitro and in vivo. In vitro experiments indicated that thermotransport within the catheter, causing a temperature measurement error, can occur. Flow model measurements were used for consequent modification of the original thermodilution formula for calculation of flow. In the in vivo investigations the thermodilution and electromagnetic methods were compared for measurement of pig portal blood flow. Using the modified formula for the flow calculations, good agreement was found between the two methods (r = 0.958). For the continuous thermodilution method in vivo the standard deviation of a single measurement was 19 ml/min and the coefficient of variation 1.6%.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
S Häggmark, B Biber, J G Sjödin, O Winsö, B Gustavsson, S Reiz. 1982. The continuous thermodilution method for measuring high blood flows.. https://pubmed.ncbi.nlm.nih.gov/7134816/