PubMed · 7210611
A method for the automated screening for hepatitis B surface antigen.
Abstract
A reverse passive hemagglutination (RPHA-G) technique for the automated screening of hepatitis B surface antigen (HBsAg) on the Groupamatic 360 (Kontron International, Zürich, Switzerland) has been developed. It is more sensitive than manual RPHA (RPHA-M) and nearly as sensitive as radioimmunoassay (RIA); it detects 6 ng/ml of HBsAg, compared to 3 ng/ml by RIA and 25 ng/ml by RPHA-M. Although it is slightly inferior to the other two methods in respect of specificity, the incidence of false-positive reactions was only 0.34%. The test can also be performed manually without the Groupamatic.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
K Kunitomo, H Saji, S Yahara, T Hosoi. 1980. A method for the automated screening for hepatitis B surface antigen.. https://doi.org/10.1111/j.1423-0410.1980.tb01860.x
Cite the original work for its findings. Save a collection to share your selection of sources.