PubMed · 7217168
Ontario accelerator dose intercomparison study.
Abstract
An ionometric dose intercomparison has been carried out on eight accelertors, with maximum photon energies from 6 to 32 meV, at five radiotherapy centres in Ontario. The ratio of the dose based on the clinically-employed rad/monitor unit to the dose measured by the committee representative had a mean value of 0.994, with a range of 7.3% and a coefficient of variation of 2.5%. The ratio of the dose measured by the institution's physicist to the dose measured by the committee representative had a mean value of 1.000 with a range of 6.9% and a coefficient of variation of 2.3%. Eight recommendations regarding dose calibration procedures are presented.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
D J Dawson, J A Rawlinson, P C Shragge. 1981. Ontario accelerator dose intercomparison study.. https://pubmed.health/record/pubmed%3A7217168/ontario-accelerator-dose-intercomparison-study
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite PubMed Health by Indiciae: BibTeX · RIS.