The humidity output of a circle system was raised to 28 mg H2O/l by the use of a modified rotating wick vaporizer placed in the center of the soda lime canister and coaxial inspiratory and expiratory limbs. Both the fresh gas inflow and the expired gases passed through the lime and reached a compartment below it. The bag/ventilator connector, bearing a pressure relief valve, opened on the lateral wall of that compartment. Gases returning to the inspiratory valve passed: (1) through a tube in the canister connecting the inferior compartment to the vaporizer above water level, (2) through the upper portion of the vaporizer and around the rotating wick, and (3) through a tube emerging from the top of the vaporizer to reach the inspiratory valve. Thus inspired gases were humidified by the rotating wick constantly replenishing its water content warmed by the reaction of neutralization. The use of coaxial inspiratory and expiratory limbs reduced water condensation outside the canister.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J Chalon, C K Tang, R Dwarakanath, R Katz, S Ramanathan, H Turndorf. 1981. A circle system with a rotating wick vaporizer.. https://doi.org/10.1007/bf03007263

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Perception of training needs and opportunities in advanced airway skills: a survey of British and Irish trainees.

BACKGROUND AND OBJECTIVE: We surveyed delegates at the Group of Anaesthetists in Training (UK) meeting to investigate evidence of a training-gap (number of fibreoptic intubations believed to bestow competence vs. number actually performed). METHODS: Questionnaires were distributed to and collected from delegates in person. Questions covered six areas, including experience of fibreoptic intubation and cricothyrotomy, fibreoptic intubation as a specialist skill and ethical issues. RESULTS: We received 221 replies (76%). All trainees believed competence to be achievable with 10 intubations (interquartile range (IQR) 10-20); the median number performed was 2 (IQR 0-4). This was statistically significant for the groups' senior house officers, 1st and 2nd year registrars and 3rd and 4th year registrars; P < 0.0001. Many final year trainees (12/20, 60%) also failed to achieve their competency target. Few trainees had seen or performed any cricothyrotomies (medians 0, IQRs 0-1 and 0-0). Most (195/208, 94%) believed that fibreoptic intubation was a core skill and 199/212 (94%) believed that all should be competent by completion of training. Ten percent (n = 208) felt it unethical to perform an awake training intubation with full consent and 10% believed it acceptable without explanation. Most (82.7%) would fibreoptically intubate an asleep patient (requiring intubation) without consent. CONCLUSION: Trainees reported a gap between their perception of competence and achievement in awake fibreoptic intubation. Simple and complex simulations and structured training programmes may help. Anaesthetists must address the ethics of clinical training in advanced airway management.

Anesthesiology↗