PubMed · 7290016
Multisegmented ion chamber for CT scanner dosimetry.
Abstract
A multisegmented, ionization chamber capable of determining dosimetric profiles from a CT scanner has been developed and tested. The chamber consists of a number of 2 mm wide electrically isolated segments from which ionization currents may be measured. Presented here are the performance characteristics of the chamber including energy response, dose linearity, and corrections for "cross talk" between segments. Sample dosimetric profiles are depicted for 3 and 6 mm nominal beam widths at two locations in a dosimetric phantom positioned in the x-ray beam of a fourth generation CT scanner. The results agree well with the conventional method of obtaining dosimetry measurements with TLD chips.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
M M Moore, R K Cacak, W R Hendee. Multisegmented ion chamber for CT scanner dosimetry.. https://doi.org/10.1118/1.595022
Cite the original work for its findings. Save a collection to share your selection of sources.