PubMed · 7360974
A simple electron cone interlock system for the clinac-18.
Abstract
A simple electron cone-collimator interlock system has been developed for the Clinac-18. Each cone is assigned a special slot in the storage cabinet and, upon being taken for treatment setup, sends out a binary coded decimal (BCD) signal to the control console. The voltage from the collimator setting must match the BCD signal or a comparator logic will drive the fault matrix to prevent the accelerator from energizing.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Y K Chan, B Walulik. 1980. A simple electron cone interlock system for the clinac-18.. https://doi.org/10.1148/radiology.135.1.7360974
Cite the original work for its findings. Save a collection to share your selection of sources.