Abstract
The MTFs of three different gamma camera systems have been measured using two methods. In one method a step function was used as the object and the calculations were made by means of a pocket calculator. This method has been compared with a conventional line source method, where a computer is needed in performing the calculations. The methods give the same results with a fairly good accuracy. Both of the methods are well suited to a regular control of the condition and functioning of a gamma camera, provided that the camera is connected to a data-processing system so that the profile curves measured can be converted into the digital form.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
T Väyrynen, U Pitkänen, K Kiviniitty. 1980. Methods for measuring the modulation transfer function of gamma camera systems.. https://doi.org/10.1007/bf00261201