Correction for deadtime losses in a gamma camera/data analysis system.
Abstract
In quantitative analyses of dynamic radionuclide studies, using large activities of 99mTc flowing through organs it is necessary to determine accurately count-rate losses and to provide appropriate corrections. A method is described for estimating deadtimes and applying corrections to the observed counting rates using a two component model where the gamma camera is considered to have a purely paralyzable deadtime and the data analysis system a purely non-paralyzable deadtime.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
K Cranley, R Millar, T K Bell. 1980. Correction for deadtime losses in a gamma camera/data analysis system.. https://doi.org/10.1007/bf00445626