Design of the EPC-9, a computer-controlled patch-clamp amplifier. 2. Software.
Abstract
The EPC-9 computer-controlled amplifier has no front-panel controls; therefore the user interface to the EPC-9 patch-clamp amplifier is defined entirely by software. This paper describes various user interfaces that have been implemented, including a high-level programming interface, a user interface based on the PostScript language, and graphical user interfaces that control the EPC-9 from data-acquisition programs. Also described are the algorithms used for automatic adjustment of the C-Fast and C-Slow transient cancellation circuitry. An overview of the procedures that perform automatic testing and calibration is given.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
F J Sigworth, H Affolter, E Neher. 1995. Design of the EPC-9, a computer-controlled patch-clamp amplifier. 2. Software.. https://doi.org/10.1016/0165-0270(94)00129-5