Output factors for irregularly shaped electron fields.
Abstract
It is necessary to know the output factors (dose per monitor unit at depth of maximum) for irregularly shaped electron beam fields to accurately deliver the prescribed dose to the target. Measuring the output factors for individually shaped electron beam fields for each patient is inconvenient. Using the measured output factors for two square fields, one can obtain the output factor for an irregular shaped electron portal with area intermediate between the areas of the two square fields, by obtaining the equivalent square area (as with photons) of the irregularly shaped field, and then interpolating between the output factors of the two square field areas to obtain the output factor for the irregularly shaped field. This empirical method offers a simple, practical solution. The accuracy of the method is about 1% to 2%, depending on the shape and size of the irregularly shaped electron field.