A patient dose verification program using diode detectors.

Abstract

In this work, we present a comprehensive patient dose verification program using a commercial dual diode detector system for external photon beam radiotherapy. To achieve acceptable accuracy, factors potentially affecting diode sensitivity need to be carefully considered in dose determination. Several authors have shown that diode sensitivity depends on temperature, saturation, linearity, radiation damage, and beam incidence angle. In addition, we found that the use of wedge filters, block trays at a close distance to the detectors, variation in radiation field size, as well as measurement SSDs may affect diode sensitivity significantly. In practice, we use a 5% dose deviation of measured dose from expected patient dose to prompt reevaluation of patient dosimetry. This yields a 90% acceptance rate in clinical patient cases.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

C Li, L S Lamel, D Tom. 1995. A patient dose verification program using diode detectors.. https://doi.org/10.1016/0958-3947(95)00018-r

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Calibration of a mosfet detection system for 6-MV in vivo dosimetry.

PURPOSE: Metal oxide semiconductor field-effect transistor (MOSFET) detectors were calibrated to perform in vivo dosimetry during 6-MV treatments, both in normal setup and total body irradiation (TBI) conditions. METHODS AND MATERIALS: MOSFET water-equivalent depth, dependence of the calibration factors (CFs) on the field sizes, MOSFET orientation, bias supply, accumulated dose, incidence angle, temperature, and spoiler-skin distance in TBI setup were investigated. MOSFET reproducibility was verified. The correlation between the water-equivalent midplane depth and the ratio of the exit MOSFET readout divided by the entrance MOSFET readout was studied. MOSFET midplane dosimetry in TBI setup was compared with thermoluminescent dosimetry in an anthropomorphic phantom. By using ionization chamber measurements, the TBI midplane dosimetry was also verified in the presence of cork as a lung substitute. RESULTS: The water-equivalent depth of the MOSFET is about 0.8 mm or 1.8 mm, depending on which sensor side faces the beam. The field size also affects this quantity; Monte Carlo simulations allow driving this behavior by changes in the contaminating electron mean energy. The CFs vary linearly as a function of the square field side, for fields ranging from 5 x 5 to 30 x 30 cm2. In TBI setup, varying the spoiler-skin distance between 5 mm and 10 cm affects the CFs within 5%. The MOSFET reproducibility is about 3% (2 SD) for the doses normally delivered to the patients. The effect of the accumulated dose on the sensor response is negligible. For beam incidence ranging from 0 degrees to 90 degrees, the MOSFET response varies within 7%. No monotonic correlation between the sensor response and the temperature is apparent. Good correlation between the water-equivalent midplane depth and the ratio of the exit MOSFET readout divided by the entrance MOSFET readout was found (the correlation coefficient is about 1). The MOSFET midplane dosimetry relevant to the anthropomorphic phantom irradiation is in agreement with TLD dosimetry within 5%. Ionization chamber and MOSFET midplane dosimetry in inhomogeneous phantoms are in agreement within 2%. CONCLUSION: MOSFET characteristics are suitable for the in vivo dosimetry relevant to 6-MV treatments, both in normal and TBI setup. The TBI midplane dosimetry using MOSFETs is valid also in the presence of the lung, which is the most critical organ, and allows verifying that calculation of the lung attenuator thicknesses based only on the density is not correct. Our MOSFET dosimetry system can be used also to determine the surface dose by using the water-equivalent depth and extrapolation methods. This procedure depends on the field size used.

Calibration

Automated high-throughput mass spectrometric analysis of synthetic oligonucleotides.

The use of automated sample preparation and data collection coupled with delayed ion extraction matrix-assisted laser desorption ionization with time-of-flight detection mass spectrometry (DE MALDI-TOF MS) enables the rapid mass analysis of oligonucleotide samples. We have incorporated a matrix-diluting and liquid-spotting automated work station and a DE MALDI mass spectrometer to provide high-throughput analysis of synthetic oligonucleotides. Automated sample preparation greatly speeds processing while the spotting function allows for precise sample placement on a 100-well sample plate. The DE MALDI MS typically determines the mass of each sample to within 0.13 +/- 0.10% of the theoretically expected mass. This mass accuracy allows confirmation of identity and detection of low-level impurities present in synthetic oligonucleotide samples with an unprecedented degree of confidence. Using this automation suite, sample preparation, spotting, and analysis of 100 samples ranging from 12 to 70 bases in length can be routinely completed in less than 90 min.

Calibration

Inter-laboratory validation of solid-phase microextraction for the determination of triazine herbicides and their degradation products at ng/l level in water samples.

The accuracy and precision of solid-phase microextraction (SPME) were validated in an inter-laboratory study including ten laboratories for the analysis of triazine herbicides and their metabolites at ng/l level in aqueous samples. The SPME conditions were optimised in order to obtain maximum sensitivity. Especially, salt addition and choice of the SPME fibre coated with Carbowax-divinylbenzene increased the sensitivity. The average detection limits were in the range from 4 to 24 ng/l for the triazine herbicides, and 20 and 40 ng/l for desisopropylatrazine and desethylatrazine, respectively. The average r2 values of the calibration curves were above 0.99 for all of the analytes. The statistical data treatment was performed in accordance with the International Standardisation Organisation (ISO) standard 5725. Relative repeatability standard deviations between 6 and 14% and relative reproducibility standard deviations between 10 and 17% were found. The determined concentrations of the reference sample compared well to the "true" values, thus proving the good accuracy of the method. It is concluded that SPME is a reliable technique for the quantitative analysis of water samples containing triazine herbicides in concentrations around the European limit of 100 ng/l for individual pesticides in drinking water.

Calibration