PubMed Health⌕ Search

PubMed · 7660150

Description of an in vitro platelet function analyzer--PFA-100.

Abstract

A new in vitro system for the detection of platelet dysfunction, PFA-100, has been developed. It provides a quantitative measure of platelet function in anticoagulated whole blood. The system comprises a microprocessor-controlled instrument and a disposable test cartridge containing a biologically active membrane. The instrument aspirates a blood sample under constant vacuum from the sample reservoir through a capillary and a microscopic aperture cut into the membrane. The membrane is coated with collagen and epinephrine or adenosine 5'-diphosphate. The presence of these biochemical stimuli, and the high shear rates generated under the standardized flow conditions, result in platelet attachment, activation, and aggregation, slowly building a stable platelet plug at the aperture. The time required to obtain full occlusion of the aperture is reported as the "closure time." We have found that impairment of von Willebrand factor, or inhibition of platelet receptors glycoprotein Ib or IIb/IIIa with monoclonal antibodies or peptides, resulted in abnormal closure times. An antifibrinogen antibody, in contrast, failed to show any effect. The test appears to be sensitive to platelet adherence and aggregation abnormalities. The PFA-100 system has potential applications in routine evaluation of platelet function in the clinical setting because of its accuracy, ease of operation, and rapid turnaround of results.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

S K Kundu, E J Heilmann, R Sio, C Garcia, R M Davidson, R A Ostgaard. 1995. Description of an in vitro platelet function analyzer--PFA-100.. https://doi.org/10.1055/s-0032-1313612. https://pubmed.health/record/pubmed%3A7660150/description-of-an-in-vitro-platelet-function-analyzer-pfa-100

Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite PubMed Health by Indiciae: BibTeX · RIS.

KEEP EXPLORING

Related citations

Blood flow velocity effects and role of activation delay time on growth and form of platelet thrombi.

Mural thrombi are composed dominantly of platelets and develop under a blood flow. Portions can break off and are carried in the blood flow as emboli. Thrombus growth rates are affected by the velocity of the blood flow, but they do not simply increase with it, they exhibit a maximum, with subsequent decrease. Whereas this variation indicates an interaction of biochemical and physical processes, studies have concentrated widely on understanding only the biochemical processes. Here we show results of simulation of thrombus formation in 3D flows by accounting for the movements of individual platelets. Each platelet follows prescribed rules for interactions while the local flow around the thrombus continuously adjusts to the growing structure of the thrombus, also when embolization occurs. With an activation delay time assigned to each platelet we demonstrate the dependence of thrombus growth rate on blood velocity as found experimentally by Begent and Born [Begent N, Born GV (1970) Nature 227:926-930]. With activated platelets having mutual tensile action sustainable up to a prescribed distance we achieve thrombus growth faster than with shorter maximum distances that make a thrombus less porous; when the prescribed maximum distance is large enough the thrombus shape is not like a "hill" but like a "carpet." We find that thrombus growth rate is enhanced by modest pulsatility but less so when pulsations are amplified in part because of more embolization.

Bleeding Time↗

Laboratory identification of von Willebrand disease: technical and scientific perspectives.

The correct diagnosis and classification of von Willebrand disease (von Willebrand disorder; vWD) is crucial because the presenting biological activity of von Willebrand factor (vWF) determines both the hemorrhagic risk and subsequent clinical management. Many laboratory assays are employed, given that assay limitations and vWD heterogeneity results in no single test being able detect all forms of vWD. Minimal laboratory identification requires assessments of vWF:antigen, factor (F) VIII:coagulant activity, and functional vWF (using vWF:ristocetin cofactor activity and vWF:collagen-binding activity). Tests to help subclassify vWD include ristocetin-induced platelet aggregation, vWF:multimers, and vWF:FVIII binding assays. New diagnostic developments are now influencing vWD diagnosis, including advancements in methodologies, automation, new platelet function analyzers, genetic mutational analysis, and a better understanding of therapeutic pharmacokinetics. This review focuses on the current recommended laboratory process for investigation of vWD from a practical scientific technical laboratory perspective. Selection of appropriate combination test panels and testing sequence is crucial for the proper diagnosis and classification of congenital vWD.

Bleeding Time↗