PubMed · 7755963
Galactose biosensors using composite polymers to prevent interferences.
Abstract
A biosensor using a composite polymer to prevent interferences was used in a flow injection analysis system for the detection of galactose in human plasma. The biosensor consisted of galactose oxidase immobilized on a platinized carbon electrode that had been modified with a composite polymer. The composite polymer showed improved selectivity to hydrogen peroxide compared with either of its individual polymeric components, Nafion and a copolymer of diaminobenzene and resorcinol. The composite polymer minimized the effect of possible interference from urate, ascorbate, and acetaminophen. This analytical system had a minimum detection limit of 50 microM, linearity to 6 mM, a storage stability of greater than 30 days, and a high sample throughput (approx. 120 samples/h).
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
P Manowitz, P W Stoecker, A M Yacynych. 1995. Galactose biosensors using composite polymers to prevent interferences.. https://doi.org/10.1016/0956-5663(95)96854-r
Cite the original work for its findings. Save a collection to share your selection of sources.