PubMed Health⌕ Search

PubMed · 7790215

Radon entry control in new house construction.

Abstract

People exposed to high concentration levels of radon face an increased risk of developing lung cancer. The risk is directly proportional to the length and level of radon exposure. Because of health reasons, it is safer to build new houses with radon mitigation systems installed in slab-on-grade houses. However, the interrelationships between parameters and factors governing radon entry and control are highly complex. A study performed by the University of Florida has examined the effectiveness of different radon entry control approaches. The analysis was based on 47 houses from three research projects conducted by the University of Florida (14 houses), Florida Solar Energy Center (13 houses), and GEOMET Technologies (20 houses). The evaluation of the performance and effectiveness of improved floor slabs, space conditioning, and ventilating systems were analyzed. Statistical analyses of the interrelationship between various parameters were also performed. Study findings such as the important factors in reducing radon entry and the effectiveness of passive construction approach and active subslab depressurization systems are presented in this paper.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

F T Najafi, L Lalwani, W G Li. 1995. Radon entry control in new house construction.. https://doi.org/10.1097/00004032-199507000-00007

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Monitoring of biofilm-associated Legionella pneumophila on different substrata in model cooling tower system.

Cooling towers have the potential to develop infectious concentrations of Legionella pneumophila. Legionella counts increases where biofilm and warm water temperatures are present. In this study, biofilm associated L. pneumophila and heterotrophic bacteria were compared in terms of material dependence. Model cooling tower system was experimentally infected by L. pneumophila standard strain and monthly monitored. Different materials were tested for a period of 180 days. The lowest L. pneumophila and heterotrophic plate counts were measured on plastic polymers, whereas L. pneumophila and heterotrophic bacteria were accumulated rapidly on galvanized steel surfaces. It can be concluded that selection of plastic polymers, as a manufacturing material, are suitable for recirculating water systems.

Air Conditioning↗

Real-time evaluation of ventilation filter-bank systems.

This study evaluated two government facility ventilation systems. One was a metropolitan government office complex with a recirculation system where outside air was the makeup air; the other was a NIOSH facility that used 100% outside air with no recirculation. The methodology employed was a modified American Society of Agricultural Engineers standard (S525) for testing total enclosure filtration efficiency, in agricultural tractor cabs, with optical particle counters (OPC). The low-efficiency bag filters were tested when new and after being in the ventilation system for 3 months. The replacement medium-efficiency filters were evaluated for 6 months (the manufacturer's suggested change-out schedule). These eight-chamber, medium-efficiency filters had an increased filter surface area that resulted in increased airflow through the system. Unfortunately, these filters contained electrostatic filter media and lost filtration efficiency rapidly, which was subsequently confirmed in a 30-day study conducted to determine an appropriate change-out schedule for the eight-chamber bag filters. The study determined that less than 6 months' use was justified due to the reduced efficiency of the electrostatic filter media. The NIOSH facility's air handler #8 (100% outside air unit) was upgraded from electrostatic bag filters, which had a suggested 9-month change-out schedule, to V Bank mechanical, wet-laid, glass fiber filters. The results of a 3-year evaluation showed that the V Bank filters had better filter efficiency after 3 years of service than the electrostatic filters had at 9 months. Both studies employed matched OPC instruments to reduce instrument-to-instrument bias. The methodology is adaptable to monitoring the total efficiency of most air filtration systems, and results can help make decisions about upgrading filter performance.

Air Conditioning↗

Integrated real-time PCR for detection and monitoring of Legionella pneumophila in water systems.

We evaluated a ready-to-use real-time quantitative Legionella pneumophila PCR assay system by testing 136 hot-water-system samples collected from 55 sites as well as 49 cooling tower samples collected from 20 different sites, in parallel with the standard culture method. The PCR assay was reproducible and suitable for routine quantification of L. pneumophila. An acceptable correlation between PCR and culture results was obtained for sanitary hot-water samples but not for cooling tower samples. We also monitored the same L. pneumophila-contaminated cooling tower for 13 months by analyzing 104 serial samples. The culture and PCR results were extremely variable over time, but the curves were similar. The differences between the PCR and culture results did not change over time and were not affected by regular biocide treatment. This ready-to-use PCR assay for L. pneumophila quantification could permit more timely disinfection of cooling towers.

Air Conditioning↗