High dose-rate induced temperature artifacts: thermometry considerations for simultaneous interstitial thermoradiotherapy.

Abstract

PURPOSE: The goal of the present study was to investigate the effect of high dose-rate radiation on a flouroptic thermometry system commonly used during microwave hyperthermia. METHODS AND MATERIALS: Measurements were performed by placing the flouroptic thermometry sensors at distances of < or = 1.5, 5, 10, and 15 mm from a remote afterloading high dose-rate 192Ir source in a water bath (at two different temperatures) and in a tissue equivalent radiation bolus medium. A simulated volumetric clinical setup using a radiation bolus medium was performed with thermometry sensors placed at 1.5, 7.5, 8.4, and 10.6 mm from a scanning high dose-rate source. RESULTS: It was found that high dose-rate radiation caused thermometry artifacts greater than 1.5 degrees C within 2 min for flouroptic thermometers placed 1.5 mm from a 5 Ci activity high dose-rate source. Simple calculations showed that artifacts of this magnitude could not be due to any heating caused by the energy deposited by the high dose-rate source. The artifact decayed, but was still evident 24 h after the exposure. The effect strongly depended on distance with a 0.7 degrees C artifactual increase in temperature seen for the probe 5 mm from the high dose-rate source. Moreover, experiments performed under conditions that represented a clinical setup with a 7 Ci high dose-rate source showed that for exposure times of 10 s, at distances of 1.5 mm, significant artifacts (> 0.5 degrees C) are produced. CONCLUSIONS: These findings indicate that high dose-rate-induced temperature artifacts should be taken into account in the quality assurance procedures for the treatment of patients with simultaneous interstitial thermoradiotherapy.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

W L Straube, A S Meigooni, E G Moros, J F Williamson, R J Myerson. 1994-09-30. High dose-rate induced temperature artifacts: thermometry considerations for simultaneous interstitial thermoradiotherapy.. https://doi.org/10.1016/0360-3016(94)90020-5

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Striping artifact removal in VisiumHD data through nuclear counts modeling.

MOTIVATION: 10x Genomics VisiumHD enables spatial transcriptomics at 2&#x2009;&#xb5;m &#xd7; 2&#x2009;&#xb5;m resolution but exhibits slide-specific, non-periodic striping artifacts due to lane-width variability. These multiplicative row/column effects distort bin total counts and can bias downstream analyses. The state-of-the-art destriping approach is the normalization procedure used as a preprocessing step in bin2cell; it applies sequential high-quantile row- then column-wise normalization, which is asymmetric and can introduce edge effects/macro-stripes and distortions of large-scale total-count structure. RESULTS: We propose a statistical destriping approach that leverages nuclei segmentation from the co-registered H&E image. Assuming transcript abundance is constant within each nucleus, we model bin counts with a negative binomial distribution whose mean is a product of a nucleus-specific concentration and row- and column-specific stripe-factors reflecting lane-width variation. We fit all parameters in a generalized linear modeling framework with cross-validated regularization on stripe-factors and iterative dispersion estimation, and use the fitted parameters to correct the observed counts into a destriped image. On synthetic data with known ground truth, our method improves stripe-factor estimation accuracy and reduces error in corrected counts relative to bin2cell and bin2cell-derived baselines. Across four public VisiumHD slides, it consistently lowers striping intensity while substantially better preserving biological signal present in the large-scale global count structure and avoiding the artifacts introduced by other methods. AVAILABILITY AND IMPLEMENTATION: All source code and links to publicly available data used for this study are available at https://github.com/paolamalsot/destriping-GLM.

Artifacts↗

Impairment of echocardiographic acoustic window caused by breast implants.

Cosmetic breast implants are increasing in popularity. The presence of a breast implant overlying the anterior mediastinal space as a cause of impairment of the echocardiographic acoustic window has not been described previously. Here, we report three cases with significant impairment of echocardiographic acoustic window caused by breast implants. Clinicians should be aware of this interference and women should be informed of this dilemma before considering this cosmetic surgery.

Artifacts↗