PubMed · 7994211
Measuring leg ulcers using a laser displacement sensor.
Abstract
A non-contact method for the measurement of a skin ulcer's area and volume has been developed. A commercially available laser displacement sensor is scanned across the ulcer's surface to produce a displacement image. From this image the healthy normal skin's surface is reconstructed and the area and volume found. Results of measurements made with phantom ulcers show that the method has a precision and accuracy +/- 5% of the total size. This compares favourably with results from other non-contact methods that have been published. The method of measurement is simple and reliable. At present, the time taken to produce a displacement image is lengthy but this is due to a limitation in the instrumentation, rather that a failing of the technique. The result of a preliminary measurement made on a patient's leg ulcer is presented.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
D A Ibbett, R E Dugdale, G C Hart, K R Vowden, P Vowden. 1994. Measuring leg ulcers using a laser displacement sensor.. https://doi.org/10.1088/0967-3334%2F15%2F3%2F011
Cite the original work for its findings. Save a collection to share your selection of sources.