PubMed · 807949
A wedge filter interlock system for 4 MV linear accelerators.
Abstract
The wedge filter interlock system on a 4 MV linear accelerator was modified to ensure proper selection and placement of filter. By using a series of microswitches to replace the two-pin connection for each wedge, proper placement of the wedge in the treatment beam was ensured. Both the chain and plug were eliminated by using a 6 pole, 6 position sub-miniature rotary switch to produce the remaining connections.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
R G Lane, K P Doppke. 1975. A wedge filter interlock system for 4 MV linear accelerators.. https://doi.org/10.1148/116.2.452