Linac-accelerator-radiosurgery.
Abstract
A survey is given of the actual possibilities and limitations of the use of linear accelerators (Linac radiosurgery systems) for intra = cranial radiosurgery. Depending on the collimator size, spherical fields from 5-54 mm in diameter can be irradiated with dose gradients from 10% (large fields) to 20% (small fields) per millimeter distance between surface and treatment volume. This is comparable to the possibilities of Gamma-Knife and Proton-irradiation. Optimal mechanical adjustment of gantry and linac table are necessary for the required stability of the isocenter. Mechanical inaccuracy should be smaller than 0.8 mm. Advanced computerized 3D-treatment planning systems are indispensable prerequisites for accurate treatment and use of the flexibility of the linac system. Future developments are outlined.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
V Sturm, W Schlegel, O Pastyr, H Treuer, J Voges, R P Müller, W J Lorenz. 1993. Linac-accelerator-radiosurgery.. https://doi.org/10.1007/978-3-7091-9297-9_20