PubMed HealthSearch

PubMed · 833181

The bone-cement interface temperature during total joint replacement.

Abstract

Sets of thermoprobe measurements of the temperature of the bone-cement interface were made during twenty total joint-replacement procedures. Although the range of rise in temperature was wide, from 3 to 17 degrees centigrade, the highest temperature obtained was 48 degrees centigrade, which is well below the denaturation point (56 degrees centigrade) of proteins. Several factors, such as the presence of blood and moisture at the interface and the large surface area and poor heat conductivity of methylmethacrylate, prevent the interface from experiencing the high rise in temperature that occurs at the center of the polymerizing cement mass.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

F W Reckling, W L Dillon. 1977. The bone-cement interface temperature during total joint replacement.. https://pubmed.ncbi.nlm.nih.gov/833181/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A finite element study of the initiation of failure of fixation in cemented femoral total hip components.

In order to study initial mechanisms of failure in cemented femoral total hip components, an anatomically accurate three-dimensional linear finite element model was constructed and verified against experimental strain measurements in the cement mantle. Good agreement was found between predicted and measured strains. The likelihood of failure initiation due to cement-prosthesis debonding and crack initiation at voids was studied for loading conditions simulating both one-legged stance and stair climbing. The "out of plane" forces involved in stair climbing appear to be the greatest threat to the fixation of total hip replacements. In stair climbing, cement-prosthesis debonding and pore crack initiation were probable in the proximal anteromedial region of the cement mantle, and near the distal tip of the implant. The proximal stresses in stair climbing were higher than the distal stresses in either stair climbing or one-legged stance.

Bone Cements

The effect of cement fixation on initial micromotion of the femoral component in condylar knee replacement.

Four fresh, cadaveric specimens were fitted with total knee replacements and mounted on a knee joint loading and measuring rig. Femoral component micromotion in the sagittal plane was measured with a transducer during flexion and extension of the specimen. Various condylar designs were used without and with cement. Movement followed a characteristic curve pattern and was less with cement fixation particularly for bone of low trabecular strength as determined by bone sample tests for stiffness. Cement fixation appears to be superior to cementless fixation.

Bone Cements