Ghost phase cancellation with phase-encoding gradient modulation.

Abstract

Motion artifacts are a dominant cause of magnetic resonance image quality degradation. Periodic or nearly periodic motion results in image replicates of the moving structures in spin-warp Fourier imaging. The replicates, or ghosts, propagate in the image in the phase-encoding, or y, direction. These ghosted images can be considered to consist of the time-averaged spin density I0 and a ghost mask g. A set of j ghosted images Ij may be acquired in which the ghost mask is intentionally phase shifted by varying amounts relative to I0 with interleaved acquisitions that have shifted phase-encoding orders or by acquiring multiple images during a single readout period in the presence of an oscillating phase-encoding gradient. The resulting complex images Ij have the same time-averaged spin density I0 but have ghost contributions gj that, on a pixel-by-pixel basis, trace part of a circle around I0. The source images Ij can then be used to estimate I0. Simulations and experiments with the phase-encoding gradient modulation method show good general ghost suppression for a variety of quasi-periodic motion sources including both respiratory-type artifacts and flow artifacts. The primary limitation of the method is the need for rapid gradient switching.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

R S Hinks, Q S Xiang, R M Henkelman. Ghost phase cancellation with phase-encoding gradient modulation.. https://doi.org/10.1002/jmri.1880030514

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Striping artifact removal in VisiumHD data through nuclear counts modeling.

MOTIVATION: 10x Genomics VisiumHD enables spatial transcriptomics at 2 µm × 2 µm resolution but exhibits slide-specific, non-periodic striping artifacts due to lane-width variability. These multiplicative row/column effects distort bin total counts and can bias downstream analyses. The state-of-the-art destriping approach is the normalization procedure used as a preprocessing step in bin2cell; it applies sequential high-quantile row- then column-wise normalization, which is asymmetric and can introduce edge effects/macro-stripes and distortions of large-scale total-count structure. RESULTS: We propose a statistical destriping approach that leverages nuclei segmentation from the co-registered H&E image. Assuming transcript abundance is constant within each nucleus, we model bin counts with a negative binomial distribution whose mean is a product of a nucleus-specific concentration and row- and column-specific stripe-factors reflecting lane-width variation. We fit all parameters in a generalized linear modeling framework with cross-validated regularization on stripe-factors and iterative dispersion estimation, and use the fitted parameters to correct the observed counts into a destriped image. On synthetic data with known ground truth, our method improves stripe-factor estimation accuracy and reduces error in corrected counts relative to bin2cell and bin2cell-derived baselines. Across four public VisiumHD slides, it consistently lowers striping intensity while substantially better preserving biological signal present in the large-scale global count structure and avoiding the artifacts introduced by other methods. AVAILABILITY AND IMPLEMENTATION: All source code and links to publicly available data used for this study are available at https://github.com/paolamalsot/destriping-GLM.

Artifacts

Motion suppression in MR imaging of the liver: comparison of respiratory-triggered and nontriggered fast spin-echo sequences.

OBJECTIVE: Our purpose was to compare the effectiveness of a respiratory-triggered fast spin-echo (RTFSE) pulse sequence and a nontriggered fast spin-echo pulse sequence for imaging liver masses. MATERIALS AND METHODS: Forty-one patients with suspected liver masses were imaged at 1.5 T with a fast spin-echo sequence and an RTFSE sequence designed to trigger data acquisition at end expiration. All other imaging parameters were identical. MR images were compared qualitatively for lesion detection and conspicuity, anatomic sharpness, vascular definition, phase artifacts, and overall image quality. Quantitative analysis was performed to obtain lesion-liver contrast and contrast-to-noise ratio (CNR) measurements of all liver masses. RESULTS: Thirty-three patients had liver masses. The RTFSE images showed superior anatomic sharpness in 83% of examinations and superior overall image quality in 85% of examinations. Lesion detection and conspicuity were superior for the RTFSE images in 53% of examinations and were similar for the two techniques in 38%. In 10 patients we detected more lesions on RTFSE images, and in one patient we detected more lesions on fast spin-echo images. In the remaining patients the number of lesions detected on RTFSE images was identical to the number detected on fast spin-echo images. Respiratory artifacts were less noticeable on the RTFSE images in 66% of examinations and on the fast spin-echo images in 14%. Quantitative analysis showed a 29% increase in mean relative lesion-liver contrast and a 34% increase in mean relative CNR for the RTFSE images. Mean lesion-liver contrast and CNR for the RTFSE images (152.6 +/- 100.9, 14.2 +/- 9.3) were superior to corresponding values for the fast spin-echo images (123.4 +/- 88.0, 10.9 +/- 7.4) (p < .0001). CONCLUSION: Compared with nontriggered fast spin-echo MR images, the RTFSE MR images were superior for our evaluation of liver masses. By acquiring data during a period of reduced respiratory motion, the RTFSE sequence produced images with sharper anatomic detail, equal or less phase ghosting, and measurable improvement in the lesion-liver contrast and CNR.

Artifacts