PubMed · 845243
Continuous aerosol therapy system using a modified Collison nebulizer.
Abstract
A Collison nebulizer was incorporated into an exposure system for administering antiviral compounds as continuous aerosols to mice infected with influenza virus. The nebulizer was modified to control aerosol output by varying the liquid feed rate. A multiple regression equation was developed from data obtained with uranine dye to define the aerosol concentration of the dye in the system as a function of the concentration of the dye in the spray fluid and the rate at which it was aerosolized. The rate of change of the concentration of the test solution due to evaporative losses was also ascertained for a 1-ml/min feed rate over a 23.5-h period of operation. Procedures are outlined for using these relationships to determine the concentration of a given drug that will result in a given dose. Performance data for the drug ribavirin are presented.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
H W Young, J W Dominik, J S Walker, E W Larson. 1977. Continuous aerosol therapy system using a modified Collison nebulizer.. https://doi.org/10.1128/jcm.5.2.131-136.1977
Cite the original work for its findings. Save a collection to share your selection of sources.