[Initial experiences using SPECT cameras with multiple detector heads].

Abstract

User acceptance testing of newly installed gamma cameras is mandatory in order to verify manufacturers' specifications and to ensure optimal image quality. Acceptance test procedures are widely standardized today and are carried out following commonly accepted test protocols. Two new triple head gamma cameras have been installed during the last year at the Department of Nuclear Medicine of the Vienna University Hospital. For acceptance testing of these cameras a modified and expanded version of the commonly used test protocol for single head SPECT systems has been employed. Details of the tests that have been carried out and a comparison of the results obtained for both cameras are presented. Furthermore, some specific problems that are encountered during acceptance testing of multiple head systems are discussed. In particular it has been noted that some important tests of intrinsic camera parameters cannot be carried out, which makes a complete verification of vendors' specifications impossible. In spite of these problems the acceptance test system for multiple head cameras presented in this work was sufficient for establishing proper working conditions for clinical use.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

A Kugi, H Bergmann. 1993. [Initial experiences using SPECT cameras with multiple detector heads].. https://pubmed.ncbi.nlm.nih.gov/8475676/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Comparison of two repositioning devices used during radiation therapy for Hodgkin's disease.

PURPOSE: Patients irradiated for Hodgkin's disease are fixed in an immobilization cradle to improve repositioning. In the early 1990s, we changed our cradle system from a "short" upper torso cradle to an extended near-total body cradle that also includes the lower torso and thighs. In this study, we assess the impact of the extended cradle on the reproducibility of patient repositioning during irradiation of Hodgkin's disease. METHODS AND MATERIALS: A total of 782 port films of 56 patients treated immediately before and after the change-over were studied to assess positioning reproducibility. Patients treated prior to 1993 were positioned in the short cradle, while those treated 1993 and later were positioned in the extended cradle. All treatment were delivered via anterior and posterior fields and treatment areas above and below the diaphragm were considered separately and together. All treatment fields were simulated and the field shape was designed on anterior and posterior radiographs. Discrepancies in field placement between the simulation radiographs and subsequent port films were noted by a radiation oncologist and requests for position adjustment (both translational and rotational shifts) were noted. The number, magnitude, and direction of any physician-requested position adjustment on port films were retrospectively reviewed. For the purpose of scoring the frequency of field misplacements, when an adjustment was noted on two port films taken during the same treatment session (i.e., a left shift on both an anterior and a posterior port film), it was scored as only one event. A two-tailed chi-square test was used to compare the differences in requested shifts in the two patient groups. RESULTS: The study population consisted of 56 patients (31 short and 25 extended cradle) representing 92 treatment sites. A total of 782 port films representing 450 treatment setups were analyzed (292 above and 158 below the diaphragm). When all port films above the diaphragm (mostly mantle fields) are considered, position adjustments were requested in 13.4% (21 out of 157) of treatment setups with the upper torso cradle and in 5.9% (8 out of 135) of treatment setups with the extended cradle (p = 0.054). When all port films below the diaphragm (mostly paraaortic/spleen and pelvic fields) are considered, position adjustments were requested in 33.8% (27 out of 80) of treatment setups with the upper torso cradle and in 16.7 % (13 out of 78) of treatment setups with the extended cradle (p = 0.056). A reduction in the frequency of both translational and rotational adjustments were seen. When both treatment sites are combined, position adjustments were requested in 20.3% (48 out of 237) of treatment setups with the upper torso cradle and in 9.9% (21 out of 213) of treatment setups when the extended cradle was used (p = 0.0086). CONCLUSIONS: The extended cradle provides superior repositioning of patients undergoing radiation therapy for Hodgkin's disease. Differences observed in setup accuracy in this study underscore the importance of aggressive immobilization of patients with Hodgkin's disease. Increased accuracy of daily setup may provide an opportunity to improve the therapeutic ratio both by increased likelihood of tumor control and decreased risk of normal tissue complications.

Equipment Design

E-speed film.

Explore the source record for details and available documents.

Equipment Design

[Dynamics of pressure and flow curves of various expiratory pressure valves].

BACKGROUND: Flow dynamics of exhalation valves in home care ventilators (CPAP-, BiLevel-, and autoadjust machines) are determined by their physical properties. The flow characteristic implies distinct curves leading to expiratory increase and inspiratory decay of pressure depending on flow rate. MATERIAL AND METHODS: For examination of pressure/flow characteristic we examined several valves (one and two circular holes, Sullivan-, Whisper-Swivel-, Silence-valve). Expiratory peak and inspiratory trough assessed at constant pressure (10 mbar) were measured with varying flow rates. RESULTS: Valves differ with their pressure characteristics up to 1.2 mbar for inspiratory and 1.0 mbar for expiratory flow at 2.0 1/s. CONCLUSION: The specific characteristic for each valve is part of the total efficiency of a home care ventilator. Their pressure flow relationship has to be taken into account for titration process and for exchange between different machines and accessories.

Equipment Design