A Cole phantom for EIT.

Abstract

A phantom was designed for testing and comparing multifrequency EIT data collection systems. The phantom simulates a cylinder of homogeneous conductor with 16 drive and 16 receive electrodes interleaved. Combinations of resistors and capacitors were used to simulate the complex impedance, Z*, of a typical tissue in the frequency range 8-2048 kHz obeying the Cole equation Z* = Z infinity + (Z0 - Z infinity)/[1 + (if/fc)(1- alpha )] where Z* is the complex impedance at frequency f, Z0 and Z infinity are the limiting values of impedance at low and high frequencies, fc is the characteristic frequency and alpha is a constant. A practical phantom was then constructed on which four different sets of spectroscopic parameters could be selected: (i) alpha = 0.20, fc = 150 kHz, Z0/Z infinity = 3.31; (ii) alpha = 0, fc = 273 kHz, Z0/Z infinity = 2.64; (iii) alpha = 0, fc = 71.1 kHz, Z0/Z infinity = 1.36; and (iv) Z0/Z infinity = 1.00 with no dispersion.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

H Griffiths. 1995. A Cole phantom for EIT.. https://doi.org/10.1088/0967-3334%2F16%2F3a%2F003

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Feasibility of neonatal lung imaging using electrical impedance tomography.

The feasibility of detecting the lungs in preterm babies using electrical impedance tomography (EIT) was investigated. A single frequency instrument using 16 electrodes to apply current and detect peripheral voltages was constructed. The instrument applied AC current of 1.5 mA peak-peak at a carrier frequency of 20 kHz. Images were reconstructed using a sensitivity-regions backprojection method. A 9-day-old preterm baby was tested and data were collected at a speed of 10 frames/s. A dynamic image showing the lungs at full inspiration referenced to expiration is illustrated in this paper. Impedance measurements taken across the chest during the first 2 s did not show a clear pattern thus demonstrating irregular breathing. Region of interest analysis were carried out on the reconstructed images and tracked with time. Fourier transforms were then performed on these signals and a fundamental frequency at 1 Hz, corresponding to normal breathing rate of 60 breaths/min, was detected. Harmonics of the signal caused distortion especially on the left lung where the effects of cardiac events were more dominant.

Electric Impedance

The phosphorus-31 spectra of dielectrophoretically reoriented tubules in the HII phase of DOPE.

31P electric field nuclear magnetic resonance measurements are described which assess the effect of electric field on the orientation of tubules comprising the HII phase of dioeleoylphosphatidylethanolamine. A model, based on dielectrophoretic effects, was used to predict that a field of 4 MV/m would change the orientation of the lipid tubules in a HII phase. The excitation pulse was biphasic to help discriminate electric field interactions with free ions or permanent dipoles from interactions with induced dipoles, as well as to control the problems of ohmic heating, electrolysis and polarisation associated with dc or unbalanced ac excitation voltages. Spectra consistent with irreversible electrorotation and with reversible and transient electrorotation were observed. No response to the electric field was seen in certain cases. The conditions for irreversible and reversible electrorotation and failure to rotate have been tabulated and are discussed. Finally, some simple models are considered, in order to calculate the energies involved, if the observed NMR spectra are interpreted as arising from lipid HII phase reorientations.

Electric Impedance

Passive electrical properties of human stratum corneum in vitro depending on time after separation.

The passive electrical properties of human skin after separation from the body are predominated by the stratum corneum. Skin within a bath medium (150 mM phosphate buffered saline) at constant temperature (37 degrees C) exhibits a characteristic change of the passive electrical properties with time. Independent of the time the locus in the Z-plane is a depressed semicircular arc. The angle between the lines from the center of the arc to the points where the locus reaches the real axis remains unchanged. The difference between the high and low frequency resistivity (R0-Rx) increases over 10 h, reaches a plateau and decays after 20 h exponentially with a time constant of about 40 h. As model for the impedance we used a 5 element electrical circuit (R0, R1, R2, C1, C2), describing 3 pathways, (0) the dc path (appendages; R0), (1) tortuous pathways around the cell structures (R1, C1) and (2) direct pathways involving the corneocytes (R2, C2). There are characteristic changes with time in the elements of the equivalent circuit up to about 200 h after excision. Dramatic changes in C1 and R2 at about this time after separation strongly suggests destruction of the lipid structures. It will be suggested that the use of separated human stratum corneum as model for in vivo yields unreliable results after this time.

Electric Impedance