Choosing a negative pressure ventilation pump: are there any important differences?

Abstract

Five negative pressure ventilator pumps were assessed to determine whether there were any differences in performance which were likely to be of clinical importance. The pumps tested were the NEV-100 (Lifecare Inc.), the Negavent Respirator DA-1 (Dima Italia), the Thompson Maxivent (Puritan Bennett), the CCP-001 and Newmarket pumps (Si-Plan Electronics Research Ltd). A patient model was employed to investigate the relationship between the pressure waveform of each pump and the tidal volume generated, the stability of performance on a continuous 8 h test, and the response to a change in leak. On the continuous tests, the stability was better than +/- 1% for the inspiratory/expiratory (I:E) ratio and rate, and better than +/- 10% for the peak negative pressure for all of the pumps. The Negavent and NEV-100 pumps generated a square wave of pressure and for the same peak negative pressure produced a tidal volume up to 30% greater than the CCP-001 and the Newmarket pumps, which produced a half sine wave. The Maxivent produced an intermediate waveform, with a plateau at peak negative pressure but a curved decay back to atmospheric pressure. It is argued that if the pressure of the pumps producing a half sine wave is made more negative to compensate and produce an equivalent tidal volume, they are more likely to induce upper airway obstruction. The leak compensation of the NEV-100 and Negavent pumps was nearly complete but took several breaths to develop, the CCP-001 and Newmarket pumps had a very rapid compensation but this was less complete, whilst the Maxivent has no mechanism for leak compensation. From these results, it seems that in patients liable to upper airway obstruction the NEV-100 and Negavent pumps would be advantageous, whilst when rapidly changing leaks impair the efficiency of negative pressure ventilation the CCP-001 and Newmarket pumps would be preferable. The Maxivent pump seems to offer no advantages.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

I E Smith, M A King, J M Shneerson. 1995. Choosing a negative pressure ventilation pump: are there any important differences?. https://doi.org/10.1183/09031936.95.08101792

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Measurement of pulsatile tidal volume, pressure amplitude, and gas flow during high-frequency percussive ventilation, with and without partial cuff deflation.

OBJECTIVE: With a high-frequency percussive ventilator and a mechanical lung model, to measure tidal volume (V(T)), pulsatile pressure amplitude (difference between peak and nadir pulsatile pressure [DeltaP];), and mean airway pressure (P (aw)) at various pulsatile frequencies, pulsatile inspiratory-expiratory ratios (I:E(p)), and pressures (measured at the interface between the pulse-generator and the endotracheal tube [P(vent)]). METHODS: With the endotracheal tube inside an artificial trachea, we manipulated the high-frequency percussive ventilation settings and adjuncts, including pulsatile frequency, I:E(p), and P(vent) by manipulating pulsatile flow. We also studied the effects of partially deflating the endotracheal tube cuff. We measured P (aw), pulsatile pressure amplitude at the carina (DeltaP(c)), and pulsatile V(T) at the carina. With the cuff partly deflated, we measured the fraction of inspired oxygen (F(IO(2))) in the gas efflux above and below the cuff. RESULTS: Increasing the pulsatile frequency from 300 cycles/min to 600 cycles/min and changing the I:E(p) from 1:3 to 1:1 significantly reduced V(T) (p < 0.001). P (aw) and DeltaP(c) were unaffected by the change in pulsatile frequency or I:E(p), except when we did not preserve the pulsatile flow. The measured V(T) range was from 19.1 mL (at 600 cycles/min) to 47.3 mL (at 300 cycles/min). Partial cuff deflation did not significantly reduce P (aw) or DeltaP(c), but it did significantly reduce V(T) and F(IO(2)). CONCLUSION: During high-frequency percussive ventilation, the pulsatile frequency is inversely related to V(T). Partial cuff deflation reduces the delivered F(IO(2)).

Air Pressure↗

Comparison of automatic and continuous positive airway pressure in a night-by-night analysis: a randomized, crossover study.

BACKGROUND: Long-term compliance is suboptimal in the treatment of the obstructive sleep apnea syndrome (OSAS). OBJECTIVES: We compared the efficacy of and the adherence to automatic continuous positive airway pressure (APAP) and constant continuous positive airway pressure (CPAP) based on a night-by-night analysis. METHODS: We performed a randomized, single-blind crossover study in 20 patients with moderate-to-severe OSAS. After diagnostic polysomnography and manual titration, patients were treated for 8 weeks with both constant CPAP and APAP in random order. Compliance and leakage were analyzed night by night using the software LOGSoft of the Magellan iPAP device. RESULTS: The reduction in the apnea/hypopnea index (baseline 32.9 +/- 19.1/h, CPAP 4.6 +/- 2.9/h, APAP 5.6 +/- 3.6/h; p < 0.001 compared to baseline) and the Epworth Sleepiness Scale (baseline 10.3 +/- 5.7, CPAP 6.6 +/- 4.8, APAP 4.9 +/- 4.6; p < 0.001 compared to baseline) did not significantly differ between the treatment modes. Leakage time and compliance per night were not statistically different (leakage CPAP 31 +/- 57 min, APAP 25 +/- 49 min; compliance CPAP 383 +/- 116 min, APAP 382 +/- 107 min). There was no correlation between leakage and compliance. Thirteen patients (65%) preferred APAP at the end of the study. CONCLUSIONS: Treatment efficacy and adherence are similar with CPAP and APAP. There is a trend towards lower leakage with APAP therapy. Patients prefer the automatic mode to fixed pressure.

Air Pressure↗