Femtomolar sensitivity using a channel-etched thin film waveguide fluoroimmunosensor.
Abstract
A dual channel, evanescent fluoroimmunoassay format is used to detect femtomolar analyte concentrations (i.e. less than 1 part per trillion [w/w]) on an etched channel siliconoxynitride thin film integrated optical waveguide. Two assays are used to demonstrate the dose-response behaviour of the sensor: (1) a direct assay of a fluorescently-labeled protein ligand binding to an immobilized protein receptor, and (2) an indirect sandwich assay of a non-fluorescent protein ligand binding to an immobilized protein receptor, as detected by the binding of a fluorescently-labeled secondary receptor protein. A red-emitting cyanine dye (Cy-5), which minimized background fluorescence and scatter losses of the waveguide, was used in both assays. To our knowledge, this is the first report of femtomolar sensitivity in an immunosensing instrument.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
T E Plowman, W M Reichert, C R Peters, H K Wang, D A Christensen, J N Herron. 1996. Femtomolar sensitivity using a channel-etched thin film waveguide fluoroimmunosensor.. https://doi.org/10.1016/0956-5663(96)83722-5