Photolithographic micromachining of silicon is a candidate technology for the construction of high-throughput DNA analysis devices. However, the development of complex silicon microfabricated systems has been hindered in part by the lack of a simple, versatile pumping method for integrating individual components. Here we describe a surface-tension-based pump able to move discrete nanoliter drops through enclosed channels using only local heating. This thermocapillary pump can accurately mix, measure, and divide drops by simple electronic control. In addition, we have constructed thermal-cycling chambers, gel electrophoresis channels, and radiolabeled DNA detectors that are compatible with the fabrication of thermocapillary pump channels. Since all of the components are made by conventional photolithographic techniques, they can be assembled into more complex integrated systems. The combination of pump and components into self-contained miniaturized devices may provide significant improvements in DNA analysis speed, portability, and cost. The potential of microfabricated systems lies in the low unit cost of silicon-based construction and in the efficient sample handling afforded by component integration.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

M A Burns, C H Mastrangelo, T S Sammarco, F P Man, J R Webster, B N Johnsons, B Foerster, D Jones, Y Fields, A R Kaiser, D T Burke. 1996-05-28. Microfabricated structures for integrated DNA analysis.. https://doi.org/10.1073/pnas.93.11.5556

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Detection of human immunodeficiency virus type 1 and type 2 antibodies by a new automated microparticle immunoassay AxSYM HIV-1/HIV-2.

A new automated microparticle enzyme immunoassay (MEIA) for the AxSYM instrument developed recently by Abbott Laboratories was compared with two established assays, i.e. HIV-1/HIV-2 3rd Gen. Plus EIA (Abbott, Delkenheim, FRG) and Wellcozyme HIV 1 + 2 (Murex Diagnostics, Dartford, England) devised for the detection of human immunodeficiency virus type 1 (HIV-1) and HIV-2 antibodies. A total of 7293 serum samples were tested by the AxSYM HIV-1/HIV-2. The test panel included seroconversions (n = 22), samples from HIV-1 and HIV-2 positive individuals, hospitalized patients, blood donors, high risk individuals. To challenge further the specificity of the assays, large numbers of EIA repeatedly reactive but Western blot negative samples, potentially cross-reactive sera, Western blot indeterminate specimens and samples from pregnant women were tested. In four seroconversion panels, HIV-1 infection was detected one bleed earlier with the AxSYM HIV-1/HIV-2 than with the Abbott HIV-1/HIV-2 3rd Gen. Plus EIA. Although the AxSYM HIV-1/HIV-2 was tested with a higher number of challenging sera than the alternative assays, the specificity was very high (99.4%). The highest number of false positive results was obtained with serum samples that were repeatedly reactive in EIAs different from those compared in the present study. The automated AxSYM system permits the testing of a large sample number in a rapid turn-around time and by random access urgent tests can be carried out even when an assay is in progress.

Automation