Compression systems are often used in hearing aids to increase the wearing comfort. A patient has to readjust frequently the gain of a linear hearing aid because of the limited dynamic hearing range and the changing acoustical conditions. A great deal of attention has been given to the static parameters but very little to the dynamic parameters. We present a general method to describe the dynamic behavior of a compression system by comparing modulations at the output with modulations at the input. The use of this method resulted in a single parameter describing the temporal characteristics of a compressor, the cut-off modulation frequency. In this paper its value is compared with known properties of running speech. A limitation of this method is the use of only small modulation depths, and the consequence of this limitation is tested. The use of this method is described for an experimental digital compressor developed by the authors, and the effects of some temporal parameters such as attack and release time are studied. This method shows the rather large effects of some of the parameters on the effectiveness of a compressor on speech. This method is also used to analyze two generally accepted compression systems in hearing aids. The theoretical method is next compared to the effects of compression on the distribution of the amplitude envelope of running speech, and it could be shown that single-channel compression systems do not reduce the distribution width of speech filtered in frequency bands. This finding questions the use of compression systems for fitting the speech banana in the dynamic hearing range of impaired listeners.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J Verschuure, A J Maas, E Stikvoort, R M de Jong, A Goedegebure, W A Dreschler. 1996. Compression and its effect on the speech signal.. https://doi.org/10.1097/00003446-199604000-00008

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

[The tasks of the ENT specialist by hearing aid fitting. Part 2: checking the fit, subjective satisfaction, ear mold and hearing aid].

The second part of this series deals with the tasks of the ENT-specialist, who must not endorse the prescription before making sure that the hearing aid recommended by the acoustician will bring sufficient hearing improvement, and that the suggested hearing aid is suitable. The report and audiograms of the acoustician must be compared to his own results, and a check of whether at least three hearing aids from different companies have been compared must be made. The primary aim of patient satisfaction can be checked by means of the "Oldenburger Inventar" (Oldenburg questionnaire), which determines the improvement of hearing ability in quiet and noise, and of auditory localization. Since the otoplastic substantially effects the amplification of the hearing aid, the ENT specialist should make sure that it does not produce feedback, that it has a second boring, and, in case of need, a Libby horn. He should also consider whether an open fitting would be more appropriate. A free-field measurement--with and without noise--shows whether the dB(opt) has been reached. The in situ measurement and the loudness rating by the Würzburger Hörfeld (Wuerzburg auditory field) are of great help in detecting fitting errors. Additional procedures, such as auditory speech training, lip reading, or technical devices, for instance FM-communication or a light alarm clock, are often employed in elderly patients.

Correction of Hearing Impairment↗

A multisite study to examine the efficacy of the otoacoustic emission/automated auditory brainstem response newborn hearing screening protocol: recommendations for policy, practice, and research.

PURPOSE: This article examines whether changes in hearing screening practices are warranted based on the results of the recent series of studies by J. L. Johnson, K. R. White, J. E. Widen, J. S. Gravel, B. R. Vohr, M. James, T. Kennalley, A. B. Maxon, L. Spivak, M. Sullivan-Mahoney, Y. Weirather, and S. Meyer (Johnson, White, Widen, Gravel, James, et al., 2005; Johnson, White, Widen, Gravel, Vohr, et al., 2005; White et al., 2005; Widen et al., 2005) that found a significant number of infants who passed an automated auditory brainstem response (A-ABR) screening after failing an initial otoacoustic emission (OAE) screening later were found to have permanent hearing loss in one or both ears. METHOD: Similar to the approach used by F. H. Bess and J. Paradise (1994), this article addresses the public health tenets that need to be in place before screening programs, or in this case, a change in screening practice (use of a 2-step screening protocol) can be justified. RESULTS: There are no data to suggest that a 2-step OAE/A-ABR screening protocol should be avoided. CONCLUSION: Research is needed before any change in public policy and practice surrounding current early hearing detection and intervention programs could be supported.

Correction of Hearing Impairment↗