Light source-induced error in computer-assisted image analysis with a video-based system.
Abstract
OBJECTIVE: We tested the hypothesis that two different light sources, an alternating current fluorescent viewbox and a direct current halogen viewbox, do not differ with respect to their ability to illuminate reproducibly a radiograph during image capture. STUDY DESIGN: Two radiographs were taken: one with four hydroxyapatite chips mounted against a dry mandible and one without the chips. They were digitally subtracted with a video-based imaging system. The procedure was repeated at different times. RESULTS: A statistically significant difference among optical density measurements was found when the alternating current fluorescent viewbox (p < 0.001) was used and was related to light intensity variation. Such effect was not observed with the direct current halogen viewbox (p = 0.873). CONCLUSION: Study design efficiency was increased by 212% with the use of the direct current halogen viewbox so that to detect a specified treatment effect with a given level of statistical confidence, the sample size has to be 2.12 times greater if the alternating current fluorescent viewbox is used.