PubMed · 8804271
Development of a telemeterized shoulder prosthesis.
Abstract
A telemeterized shoulder prosthesis was developed and tested, first in vitro and then in a cadaver implantation. The prototype prosthesis was instrumented with 3 strain gauges fixed within a hollow, tapered neck element whose output was used to determine the load vectors experimentally applied to the humeral head. Testing demonstrated accuracy of force measurement resolution to within 5% of actual applied forces. This design seems feasible for further investigations to develop a totally implantable telemeterized prosthesis.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
F J Kummer, T R Lyon, J D Zuckerman. 1996. Development of a telemeterized shoulder prosthesis.. https://doi.org/10.1097/00003086-199609000-00004