PubMed HealthSearch

PubMed · 8829940

Five oxygen-nitrous oxide proportioning systems compared.

Abstract

The majority of contemporary gas delivery modules, on anaesthetic workstations, are equipped with oxygen-nitrous oxide proportioning systems which should prevent the delivery of hypoxic gas mixtures. We investigated five modules, of two different types, using fresh gas flows ranging from 50 mL min-1 to 20 L min-1 with minimally acceptable proportions of oxygen and maximally acceptable proportions of nitrous oxide set at the flow control valves. The oxygen concentrations of the resulting gas mixtures were measured at the fresh gas outlet of the gas delivery modules. All the modules prevented the delivery of hypoxic fresh gas flows both under normal and abnormal working conditions (unequal supply of gas pressures). All systems showed increased O2 concentrations at fresh gas flows below 1 L min-1. The systems can be used with low flow anaesthesia techniques with one exception (Dameca) also with minimal flow techniques.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

P Feigenwinter, T H Pohle, A M Zbinden. 1996. Five oxygen-nitrous oxide proportioning systems compared.. https://doi.org/10.1097/00003643-199601000-00013

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

An early approach to controlled hypotension: Felice Visalli and the "pneumoanemizator.".

In this historical vignette, the authors describe the conception and realization of the "pneumoanemizator," an ingenious pneumatic apparatus originally designed in the 1940s for the purpose of controlled hypotension. The authors also give a biographical sketch of its inventor, the pioneering Italian neurosurgeon Felice Visalli. The general working features of this nearly forgotten device and its performance in experimental and clinical trials are evoked in the light of a brief history of controlled hypotension.

Anesthesiology