Statistical study of a gamma camera transfer function.

Abstract

In order to investigate the theoretical resolution of a gamma camera, the position transfer function of the input converters, scintillator and photocathode, has been studied. It is assumed that point scintillations occur in a scintillator at xy0 at a height h above a plane photocathode of sensitivity p per incident photon and that the spatial emission pattern of photoelectrons is known completely. A statistical analysis is given of various methods for determination of the estimator x0 of x0 both for a photocathode of finite and infinite extent. It is shown that the standard deviation sigma of this estimator x0 has a minimum value 2h/(N0p)1/2 where N0p/2 is the total number of photoelectrons emitted by the cathode.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

S Degerine, M Laval. 1977. Statistical study of a gamma camera transfer function.. https://doi.org/10.1088/0031-9155%2F22%2F4%2F009

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Effects of a reduction in crystal thickness on Anger-camera performance.

The performances of two commerical gamma cameras, with crystals 1.3 and 0.6 cm thick, are compared. Phantom studies simulating clinical conditions showed no significant difference in performance at 140 keV. At 68 keV, the thinner crystal gave marginal improvements in camera performance with phantoms simulating clinical conditions. Frequent use with very low-energy emitters, such as Tl-201, would be needed to justify the expense required to refit the 1.3 cm camera with a 0.6 cm crystal.

Radionuclide Imaging