Three-dimensional strain-rate imaging.

Abstract

Strain-rate imaging uses large velocity encoding gradients to obtain measurements of velocity that are extremely insensitive to the effects of random noise. The spatial differential of velocity yields the velocity gradient from which the strain-rate and twist-rate tensors can be determined. These tensors represent the distortion of the material and are of interest in the analysis of the dynamic behavior of living tissue (e.g., that of the myocardium). This work presents a new technique that uses the magnitude of the signal in the velocity encoded data to measure through-plane velocity variations at the resolution of the voxel size. The magnitude of the MR signal contains information about the range of phases present within a voxel. When the phase is dependent on the velocity (as in phase velocity imaging), the magnitude contains information about the range of velocities within a voxel. The method presented in this work uses unbalanced slice-refocusing gradients to sample the magnitude variation introduced by the interaction of velocity encoding gradients with spatially dependent velocities. The previously developed in-plane velocity gradient methods can be easily integrated with this new through-plane measurement to characterize the deformation of the myocardium in three spatial dimensions with high accuracy. The applicability of these methods is demonstrated theoretically, in phantoms and in vivo.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

M D Robson, R T Constable. 1996. Three-dimensional strain-rate imaging.. https://doi.org/10.1002/mrm.1910360406

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Selection of high-definition 2D virtual profiles with multiple RF pulse excitations along interleaved echo-planar k-space trajectories.

A method for spatially selective excitation of 2D RF profiles is reported. The method makes use of multiple shots to traverse interleaved echo-planar trajectories in 2D k space during each RF pulse excitation. Results from each of the interleaved excitations are summed, with the net effect being the excitation of a virtual profile. The method allows for the excitation of high-definition 2D profiles with standard gradient hardware. Signal to noise is enhanced by a factor equal to the square root of the number of interleaved excitations, compared with a single-shot excitation. Potential applications for volume-localized spectroscopy, functional MRI, and high-resolution reduced-field-of-view imaging are discussed.

Echo-Planar Imaging

Multi-echo segmented k-space imaging: an optimized hybrid sequence for ultrafast cardiac imaging.

Cardiac magnetic resonance imaging requires high temporal resolution to resolve motion and contrast uptake with low total scan times to avoid breathing artifacts. While spoiled gradient echo (SPGR) imaging is robust and reproducible, it is relatively inefficient and requires long breath-holds to acquire high time resolution movies of the heart. Echo planar imaging (EPI) is highly efficient with excellent signal-to-noise ratio (SNR) behavior; however, it is particularly difficult to use in the heart because of its sensitivity to chemical shift, susceptibility, and motion. EPI may also require reference scans, which are used to measure hardware delays and phase offsets that cause ghosting artifacts; these reference scans are more difficult and less reliable in the heart. Consequently, a hybrid EPI/SPGR sequence is proposed for application to rapid cardiac imaging. A detailed optimization of SNR and echo train length for multi-echo sequences is presented. It is shown that significant reductions in total scan time are possible while maintaining good image quality. This will allow complete motion sampling of the entire heart in one to three breath-holds, necessary for MR cardiac dobutamine stress testing. Improved speed performance also permits sampling of three to six slices every heartbeat for bolus injection perfusion studies.

Echo-Planar Imaging

Further evaluation of the initial negative response in functional magnetic resonance imaging.

The initial negative response (i.e., the dip) in the functional MR signal at stimulus onset has aroused much interest. Such a response is consistent with optical imaging data and can be potentially mapped to generate spatially more specific maps. However, there are still controversies regarding the exact origin of the initial response. In particular, experimental reports of its echo-time dependence have been inconsistent. Furthermore, several investigators have suggested the possibility of an apparent dip that may arise artifactually when the interstimulus interval (ISI) is not sufficiently long. The present study investigates the echo-time dependence of the initial response and the effect of the ISI on the initial response. Experimental results obtained at TEs of 21, 30, and 45 msec demonstrate that the initial dip has a TE dependence that is in agreement with a T2* contrast, and thereby consistent with a blood oxygenation level-dependent origin. At an ISI of 90 sec, a statistically significant initial negative response was detected and shown to be indistinguishable from that observed at an ISI of 45 sec, which was used in our previous studies, indicating that the initial negative response observed at 4 T is not a consequence of short ISI.

Echo-Planar Imaging