PubMed · 8897366
Circular sampling: perspective of a time-saving scanning procedure.
Abstract
A time-saving pulse sequence that requires relatively low gradients and slow rise time is suggested herein. With this pulse sequence the entire k domain is sampled along a set of concentric circles using only 50% of the number of pulses required by the standard line sampling technique (2DFT). This can result in saving about half the total scanning time in gated cardiac imaging. The sampling configuration in the k domain is identical to MR projection reconstruction, with the advantage that a considerably smaller number of excitations is required. Furthermore, computer simulations indicate that the proposed method is also advantageous over the "half-Fourier" method, as it provides better SNR for equal acquisition time (0.83 vs. 0.707 of the 2DFT SNR respectively.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
H Azhari, O E Denisova, A Montag, E P Shapiro. 1996. Circular sampling: perspective of a time-saving scanning procedure.. https://doi.org/10.1016/0730-725x(96)00089-6. https://pubmed.health/record/pubmed%3A8897366/circular-sampling-perspective-of-a-time-saving-scanning-procedure
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite PubMed Health by Indiciae: BibTeX · RIS.