PubMed HealthSearch

PubMed · 8914873

Current-based transthoracic defibrillation.

Abstract

This study examines in a prospective, multicenter trial the feasibility and advantage of current-based, transthoracic defibrillation. Current-based, damped, sinusoidal waveform shocks of 18, 25, 30, 35, or 40 amperes (A) were administered beginning with 25 A for polymorphic ventricular tachycardia (VT) and ventricular fibrillation (VF) or 18 A for monomorphic VT; success rates were compared with those of energy-based shocks beginning at 200 J for VF/polymorphic VT and 100 J for VT. The current-based shocks were delivered from custom-modified defibrillators that determined impedance in advance of any shock using a "test-pulse" technique; the capacitor then charged to the exact energy necessary to deliver the operator-selected current against the impedance determined by the defibrillator. Three hundred sixty-two patients received > 1 shock for VF, polymorphic VT, or monomorphic VT: 569 current- based shocks and 420 energy-based shocks. Current-based shocks of 35/40 A achieved success rates of up to 74% for VF/polymorphic VT; 30 A shocks terminated 88% of monomorphic VT episodes. Energy-based shocks of 300 J terminated 72% of VF/polymorphic VT; 200-J shocks terminated 89% of monomorphic VT. We could not demonstrate a significant increase in the success rate of current-based shocks over energy-based shocks for patients with high transthoracic impedance; this may be due to inadequate sample size. Thus, current-based defibrillation is clinically feasible and effective. A larger study will be needed to test whether current-based defibrillation is superior to energy-based defibrillation.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

R E Kerber, R A Kieso, M G Kienzle, B Olshansky, A L Waldo, M D Carlson, D J Wilber, A M Aschoff, S Birger, F Charbonnier. 1996-11-15. Current-based transthoracic defibrillation.. https://doi.org/10.1016/s0002-9149(96)90062-4

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

ECG diagnosis of native heart ventricular tachycardia in a heterotopic heart transplant recipient.

A case is reported of haemodynamic collapse in a 51 year old male heterotopic heart transplant recipient caused by native heart ventricular tachycardia. An accurate diagnosis was made by selective right and left sided electrocardiography. Synchronised electrical cardioversion of the native heart (200 J) resulted in restoration of sinus rhythm with prompt relief of symptoms and amelioration of the clinical situation.

Electric Countershock

Public locations of cardiac arrest. Implications for public access defibrillation.

BACKGROUND: The purpose of this study was to describe the public locations of cardiac arrest and to estimate the annual incidence of cardiac arrest per site to determine optimal placement of automatic external defibrillators (AEDs). This was a retrospective cohort study. METHODS AND RESULTS: Locations of cardiac arrest were abstracted from data collected by emergency medical service programs in Seattle and King County, Washington, from January 1, 1990, through December 31, 1994. Types of commercial and civic establishments were tallied and grouped into 23 location categories consistent with Standard Industrial Codes, and the number of sites within each location category was determined. With the addition of "public outdoors" and "automobiles" as categories, there were 25 location categories. During the study period, 7185 arrests occurred, 1130 (16%) of which were in public locations. An annual incidence of cardiac arrest per site was calculated. Ten location categories with 172 sites were identified as having a higher incidence of cardiac arrest (> or = .03 per year per site). Thirteen location categories had a lower incidence of arrest (< or = .01 per year per site). There were approximately 71,000 sites within these categories. CONCLUSIONS: Placement of 276 AEDs in the 172 higher-incidence sites would have provided treatment for 134 cardiac arrest patients in a 5-year period, 60% of whom were in ventricular fibrillation. We estimate between 8 and 32 lives could be saved in 5 years. To cover the remaining 347 arrests occurring in public in a 5-year period, defibrillators would have to be placed in 71,000 sites, not including outdoors and automobiles. Placement of AEDs in public locations can be guided by the site-specific incidence of arrest.

Electric Countershock