PubMed · 912882
Hemoglobin interference with Du Pont Automatic Clinical Analyzer procedure for calcium.
Abstract
Hemoglobin in concentrations of approximately 2.5 g/liter or greater significantly increases values for apparent calcium in serum as determined with the Du Pont aca, an interference attributable to the 575-nm absorption band of oxyhemoglobin being uncompensated by the 577--600 nm differential absorption reading of the calcium o-cresolphthalein complexone complex.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
W H Porter, J R Carroll, R E Roberts. 1977. Hemoglobin interference with Du Pont Automatic Clinical Analyzer procedure for calcium.. https://pubmed.ncbi.nlm.nih.gov/912882/
Cite the original work for its findings. Save a collection to share your selection of sources.