Pressure vs flow triggering during pressure support ventilation.

Abstract

BACKGROUND: Adult mechanical ventilators have traditionally been pressure- or time-triggered. More recently, flow triggering has become available and some adult ventilators allow the choice between pressure or flow triggering. Prior studies have supported the superiority of flow triggering during continuous positive airway pressure, but few have compared pressure and flow triggering during pressure support ventilation (PSV). The purpose of this study was to compare pressure and flow triggering during PSV in adult mechanically ventilated patients. METHODS: The study population consisted of 10 adult patients ventilated with a mechanical ventilator (Nellcor-Puritan-Bennett 7200ae) in the PSV mode. In random order, we compared pressure triggering of -0.5 H2O, pressure triggering -1 cm H2O, flow triggering of 5/2 L/min, and flow triggering 10/3 L/min. Pressure was measured for 5 min at the proximal endotracheal tube using a data acquisition rate of 100 Hz. From the airway pressure signal, trigger pressure (deltaP) was defined as the difference between positive end-expiratory pressure (PEEP) and the maximum negative deflection prior to onset of the triggered breath. Pressure-time product (PTP) was defined as the area produced by the pressure waveform below PEEP during onset of the triggered breath. Trigger time (deltaT) was defined as the time interval below PEEP during onset of the triggered breath. RESULTS: A pressure trigger of -0.5 cm H2O was significantly more sensitive than the other trigger methods for deltaP, PTP, and deltaT (p<0.001). There was also a significant difference between patients for deltaP, deltaT, and PTP for each trigger method (p<0.001). CONCLUSIONS: For this group of patients, flow triggering was not superior to pressure triggering at -0.5 cm H2O during PSV.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

R Goulet, D Hess, R M Kacmarek. 1997. Pressure vs flow triggering during pressure support ventilation.. https://doi.org/10.1378/chest.111.6.1649

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

The effect of acute pain crisis on exhaled nitric oxide levels in children with sickle cell disease.

Exhaled nitric oxide (FE(NO)) has been shown to be decreased in children with sickle cell disease. We sought to evaluate the effect of sickle cell vaso-occlusive crisis (VOC) on FE(NO) levels. We measured FE(NO) levels in 42 children with sickle cell disease, 29 in their baseline health and 13 during an acute VOC. There was no difference in FE(NO) levels between children at baseline (15.12 +/- 9.32 ppb) and those during an acute VOC (15.68 +/- 7.26 ppb; P = 0.794). FE(NO) is not a useful marker of acute VOC in children with sickle cell disease.

Acute Disease↗

[Megadolichobasilar anomaly causing acute deafness with vertigo].

Megadolichobasilar anomaly, a dilatant arteriopathy of the basilar artery attributable to chronic arterial hypertension, can cause cranial nerve compression syndromes of the cerebellopontine angle or infarcts of the vertebrobasilar circulation. In this paper, we report on a patient with known megadolichobasilar anomaly and a partially thrombosed fusiform aneurysm of the basilar artery, who presented with acute-onset vertigo and subsequent deafness due to thromboembolic occlusion of the labyrinthine artery. Because of the vascular origin of the patient's symptoms, his vertigo disappeared over time while the deafness persisted.

Acute Disease↗

Retrieval of a leaflet escaped in a Tri-technologies bileaflet mechanical prosthetic valve.

The escape of the prosthetic heart valve disc is one of the causes of prosthetic dysfunction that requires emergency surgery. The removal of the embolized disc should be carried out because of the risk of a progressive extrusion on the aortic wall. Several imaging techniques can be used for the detection of the missing disc localization. In this report we describe a 32-year-old man who underwent mitral valve replacement with a Tri-technologies bileaflet valve three years ago, and was admitted in cardiogenic shock. Transesophageal echocardiography showed acute-onset massive mitral regurgitation. The patient underwent emergency replacement of the prosthetic valve. Only one of the two leaflets remained in the removed prosthetic valve. The missing leaflet could not be found within the cardiac cavity. The abdominal fluoroscopic study and plain radiography were unable to detect the escaped leaflet. The abdominal computed tomography scan and the ultrasound showed the escaped leaflet in the terminal portion of the aortic bifurcation. To retrieve the embolized disc laparotomy and aortotomy were performed three months later. The escaped leaflet shows a fracture of one of the pivot systems caused by structural failure. This kind of failure mode is usually the result of high stress concentration.

Acute Disease↗