PubMed · 9232172
Adjustable temporary venous spring filter: in vitro assessment.
Abstract
RATIONALE AND OBJECTIVES: The authors performed an in vitro study to evaluate a temporary venous spring filter that can extend vessels of a wide range of diameters to a slit-shaped canal. MATERIALS AND METHODS: Filters were placed in thin polyethylene tubes or in porcine inferior vena cava specimens (diameter, 10.0-16.0 mm). Physiologic saline was substituted for blood flow, and blood clots of three sizes (6 x 10, 6 x 20, and 9 x 20 mm) were funneled to the filter. Clot-trapping ability was assessed by the degree of luminal extension. RESULTS: When the luminal extension was increased from 2.6r to 2.8r (where r is the original radius of the vein models), clot-trapping ability increased significantly (61.1%-87.5% in polyethylene tubes and 15.8%-77.5% in venous specimens, P < .001). When the luminal extension was increased to 3.0r, more than 90% of the clots were trapped in all tubes. After trapping the small, medium, and large clots, mean intraluminal pressure elevation was 1.0, 1.0, and 17.0 cm of saline, respectively. CONCLUSION: Despite its simple design, the spring filter proved to be an efficient filtering device.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
J Hosaka, S Roy, K Kvernebo, I Enge, F Laerum. 1997. Adjustable temporary venous spring filter: in vitro assessment.. https://doi.org/10.1016/s1076-6332(97)80239-5