Absorption-edge fluoroscopy using a three-spectrum technique.

Abstract

We have recently reported on a 1-kVp, two-filter image subtraction method for visualizing low concentrations of elements like iodine which have K-shell absorption edges in the diagnostic x-ray energy range. However, in the application of this technique to human thyroid imaging, superimposed images due to variations in tissue and bone thickness presented serious difficulties. In this paper, a technique is described for implementing a 3-kVp, three-filter approach. Using carefully chosen spectra and logarithmic image processing, images are produced which are compatible with our previously described two-stage storage-tube subtraction device. Proper manipulation of the resulting difference images results in a final absorption-edge image in which only the element of interest is visualized, with unwanted background images suppressed. Computer calculations are presented to illustrate the relative sizes of difference signals arising from the element of interest and from tissue and bone backgrounds. Phantom studies using iodine concentrations as small as 1 mg/cm2, with variations of 10 cm of tissue and 2 g/cm2 of bone, suggest that the theory is sound and that, with straightforward apparatus modifications, images of good quality should be possible.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

F Kelcz. Absorption-edge fluoroscopy using a three-spectrum technique.. https://doi.org/10.1118/1.594219

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Analysis of laryngeal movement by digital subtraction laryngography.

Fine movement of laryngeal components was analyzed by digital subtraction laryngography with the use of a digital fluorography system. Two kinds of image processing modes, the super pulse image subtraction mode and the time interval difference mode, were used for this purpose. By the super pulse image mode, the distance and extent of movement of laryngeal components at vocal initiation and during phonation could be observed. By the time interval difference mode, the amount and direction of motion for a certain interval and fine movement, such as vocal initiation, were determined.

Fluoroscopy

A figure of merit for the assessment of image intensifier systems.

A dimensionless figure of merit (F), which combines key aspects of the performance of image intensifier systems, is derived. This figure is ideally independent of input air kerma rate and field size and relates to the fundamental characteristics of the image reception chain. In practice it is calculated from measurements of the threshold contrast (noise) and limiting resolution using the Leeds test objects and the input air kerma rate. Several practical examples illustrate its usefulness in simple performance assessment and quality assurance procedures. A comparison is made with the use of detection index diagrams.

Fluoroscopy